Chapter 8 197
Performance Tests
AF Generator AC Level Accuracy Performance Test 8
AF Generator AC Level Accuracy
Performance Test 8
The purpose of this test is to verify that the Test Set meets the
specification limits in PTR (Performance Test Record) Table 9-8, “AF
Generator AC Level Accuracy Test 8 Record,” on page 256. There are
two audio generators. AC level accuracy is measured directly with a
digital multi meter.
Figure 8-13 Setup
Procedure
1. Set the multimeter to measure AC volts.
2. On the Test Set:
a. Press
Preset.
b. Select the RF GENERATOR screen.
c. Set the AFGen1 To and AFGen2 To fields to Audio Out.
3. On the Test Set for Audio Frequency Generator 1 do the following:
a. Set the AFGen2 To level field to Off.
b. Set the audio frequency and level as shown in the Performance
Test Record (PTR) and measure the AC level. Compare the
measured voltage to the limits.
4. On the Test Set for Audio Frequency Generator 2 do the following:
a. Set the AFGen1 To level field to Off and AFGen2 To level field to
On.
b. Set the audio frequency and level as shown in the PTR and
measure the AC level. Compare the measured voltage to the
limits.
INPUT
AUDIO OUT
Multimeter
Komentáře k této Příručce