Performance Tests Power Tests
104 Agilent 81600B Tunable Laser Source Family, Fourth Edition
12
Determine the maximum deviation and the minimum deviation from
REF and record them in the Test Record.
13 Subtract the minimum deviation from the maximum deviation and
record the result as the Flatness.
Test Procedure - High Power output, using the built-in attenuator
This test procedure is applicable to the:
It measures the power flatness of the module at the outputs indicated and
when the built-in attenuator is used.
1 Set up the equipment as shown in Figure 31:
Figure 31 Test Setup for Power Flatness Tests - High Power output, using the built-in
attenuator
Agilent 81600B #200 (Output 2, High Power)
Agilent 81600B #160 (Output 2, High Power)
Agilent 81600B #150 (Output 2, High Power)
Agilent 81600B #140 (Output 2, High Power)
Agilent 81600B #142 #003
Agilent 81600B #130 (Output 2, High Power)
Zobrazit stránku 103