Agilent B2200 Users Guide, Edition 2 5-5
SCPI Command Reference
*CLS
*CLS
This command clears the Status Byte Register, the Standard Event Status Register,
and the Error Queue. This command does not clear the enable registers. See “Status
Reporting Structure” on page 5-53.
Also, this command stops the monitoring of pending operations by the *OPC
command.
This command does not have query form.
Syntax *CLS
Example OUTPUT @Agb2200;"*CLS"
*ESE
This command sets the bits of the Standard Event Status “Enable” Register.
1 enables, 0 masks.
Syntax *ESE enable_number
Query response enable_number <newline><^END>
enable_number is decimal integer value that is the sum of the binary-weighted
values of the “Enable” register bits.
Semantics The Standard Event Status “Enable” Register determines which bits of the Standard
Event Status Register are enabled. Enabled bits are ORed together, and the result is
reported to bit5 of the Status Byte Register.
The Standard Event Status “Enable” Register consists of 16 bits, but only the lower
8 bits are used, which correspond to the bits of the Standard Event Status Register.
For details, see “Standard Event Status Enable Register” on page 5-59.
Parameter Explanation
enable_number decimal integer (that is the sum of the binary-weighted values
for the desired bits), hexadecimal, octal, or binary value