Agilent TechnologiesAgilent B2200AFemto Leakage Switch MainframeAgilent B2201A14ch Low Leakage Switch MainframeUser’s Guide
3-38 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusGPIB Displays the sub menu that provides the following functions.• EXECUTEDi
4 Programming
4-2 Agilent B2200 User’s Guide, Edition 4ProgrammingThis chapter describes the automatic control programming of the Agilent B2200.•“Programming Basics
Agilent B2200 User’s Guide, Edition 4 4-3ProgrammingProgramming BasicsProgramming BasicsThis section explains the commands used for the fundamental sw
4-4 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming BasicsFundamental CommandsThe following commands are used to set the fundamental switc
Agilent B2200 User’s Guide, Edition 4 4-5ProgrammingProgramming BasicsSwitch Control The following commands are used to control open/close of the spec
4-6 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming BasicsIn the channel_list, you can specify multiple channels by using comma (,) or col
Agilent B2200 User’s Guide, Edition 4 4-7ProgrammingProgramming ExamplesProgramming Examples This section provides examples of control programs for th
4-8 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesConnecting Input-Output PathsThe following example connects instrument output
Agilent B2200 User’s Guide, Edition 4 4-9ProgrammingProgramming ExamplesTable 4-3 Input-Output Connection ExampleImports Ivi.visa.interop '1Modul
ContentsAgilent B2200 User’s Guide, Edition 41. Introduction Agilent B2200 Series . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-10 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesUsing Bias Mode The following example uses the bias mode.Figure 4-3 Bias Mode
Agilent B2200 User’s Guide, Edition 4 4-11ProgrammingProgramming ExamplesTable 4-4 Bias Mode ExampleImports Ivi.visa.interop '1Module Module1Sub
4-12 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesUsing Ground Mode The following example uses the ground mode.Figure 4-4 Groun
Agilent B2200 User’s Guide, Edition 4 4-13ProgrammingProgramming ExamplesTable 4-5 Ground Mode ExampleImports Ivi.visa.interop '1Module Module1Su
4-14 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesUsing Couple ModeThe following example uses the couple mode.Figure 4-5 Couple
Agilent B2200 User’s Guide, Edition 4 4-15ProgrammingProgramming ExamplesTable 4-6 Couple Mode ExampleImports Ivi.visa.interop '1Module Module1Su
4-16 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesSaving Input/Output LabelsYou can define labels to use for the input/output p
Agilent B2200 User’s Guide, Edition 4 4-17ProgrammingProgramming ExamplesTable 4-7 Label Definition and Data Save ExampleImports Ivi.visa.interop &apo
4-18 Agilent B2200 User’s Guide, Edition 4ProgrammingProgramming ExamplesDefining Comment for Internal MemoryYou can define comments for the internal
Agilent B2200 User’s Guide, Edition 4 4-19ProgrammingProgramming ExamplesTable 4-8 Memory Comment Definition ExampleImports Ivi.visa.interop '1Mo
Agilent B2200 User’s Guide, Edition 4ContentsSelf-Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
4-20 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationCapacitance CompensationWhen the capacitance/conductance measurement is p
Agilent B2200 User’s Guide, Edition 4 4-21ProgrammingCapacitance CompensationRequired ConditionsThe following conditions must be satisfied to use the
4-22 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationIn Figure 4-6, C2H, C2L, C3H, C3L are the compensation coefficients defin
Agilent B2200 User’s Guide, Edition 4 4-23ProgrammingCapacitance CompensationTo Create Compensation Data FileThis section explains how to create the c
4-24 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationTable 4-9 Template Compensation Data FilesFile nameaa. <path>: driv
Agilent B2200 User’s Guide, Edition 4 4-25ProgrammingCapacitance CompensationTable 4-10 Compensation Coefficients and Modifications2. Measure the R, L
4-26 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationTo obtain compensation coefficientsObtain the compensation coefficients a
Agilent B2200 User’s Guide, Edition 4 4-27ProgrammingCapacitance CompensationTo Perform Measurement and CompensationPerform the capacitance measuremen
4-28 Agilent B2200 User’s Guide, Edition 4ProgrammingCapacitance CompensationTable 4-13 Capacitance Compensation Program Example#include "stdaf
5 SCPI Command Reference
ContentsAgilent B2200 User’s Guide, Edition 4To Set GPIB Address . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-2 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceThis chapter describes the following for Agilent B2200:• SCPI commands available to con
Agilent B2200 User’s Guide, Edition 4 5-3SCPI Command ReferenceTextual Notation for Subsystem Commands CAPITAL LETTERS Capital letters are the minima
5-4 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommands SummaryCommon Commands This section describes common commands and queries, whi
Agilent B2200 User’s Guide, Edition 4 5-5SCPI Command Reference*CLS*CLS This command clears the Status Byte Register, the Standard Event Status Regist
5-6 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference*ESR?The following table shows the bits of the Standard Event Status Register and binar
Agilent B2200 User’s Guide, Edition 4 5-7SCPI Command Reference*IDN?Semantics The following table shows the bits of the Standard Event Status Registe
5-8 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference*OPC*OPC This command starts to monitor pending operations, and sets/clears the Operati
Agilent B2200 User’s Guide, Edition 4 5-9SCPI Command Reference*RST*RST This command performs an instrument reset. Status after *RST is shown below:Ch
5-10 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference*SRE*SRE This command sets the Service Request “Enable” Register bits. 1 enables, 0 ma
Agilent B2200 User’s Guide, Edition 4 5-11SCPI Command Reference*STB?The following is example for query:OUTPUT @Agb2200;"*SRE?" ENTER @Agb22
Agilent B2200 User’s Guide, Edition 4ContentsUsing Couple Mode. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-12 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference*TST?*TST? This query command executes an internal self-test, then returns the result.
Agilent B2200 User’s Guide, Edition 4 5-13SCPI Command ReferenceCommand SummarySubsystem CommandsCommand SummaryROUT subsystem :ROUTe subsystem has co
5-14 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommand Summary[:ROUT]:SYMB:CHAN card_number,channel,'string'[:ROUT]:SYMB:CH
Agilent B2200 User’s Guide, Edition 4 5-15SCPI Command ReferenceCommand SummaryCommand Description Bias Mode Commands[:ROUT]:BIAS:CHAN:DIS:CARD card_n
5-16 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommand SummaryGround Mode Commands[:ROUT]:AGND:CHAN:DIS:CARD card_number Ground-disab
Agilent B2200 User’s Guide, Edition 4 5-17SCPI Command ReferenceCommand Summary[:ROUT]:AGND:UNUSED card_number,'enable_port'[:ROUT]:AGND:UNU
5-18 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommand SummaryDIAG subsystem :DIAGnostic subsystem has commands for executing the sel
Agilent B2200 User’s Guide, Edition 4 5-19SCPI Command ReferenceCommand SummarySYSTEM subsystem:SYSTem subsystem is a collection of functions that are
5-20 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceCommand Summary:SYST:MEMO:COMM memory_number,'comment':SYST:MEMO:COMM? memor
Agilent B2200 User’s Guide, Edition 4 5-21SCPI Command Reference:DIAGnostic:TEST:CARD:CLEar:DIAGnostic:TEST:CARD:CLEar This command clears the relay t
ContentsAgilent B2200 User’s Guide, Edition 4[:ROUTe]:AGND:CHANnel:DISable[:LIST] . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5-24
5-22 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:DIAGnostic:TEST:FRAMe:CLEarSyntax :DIAGnostic:TEST:CARD:STATe? card_numberQuery resp
Agilent B2200 User’s Guide, Edition 4 5-23SCPI Command Reference:DIAGnostic:TEST:FRAMe:STATe?Syntax :DIAGnostic:TEST:FRAMe[:EXECute]? CONTroller | FP
5-24 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:AGND:CHANnel:DISable:CARDExample OUTPUT @Agb2200;":DIAG:TEST:FRAM:STAT?
Agilent B2200 User’s Guide, Edition 4 5-25SCPI Command Reference[:ROUTe]:AGND:CHANnel:ENABle:CARDExample OUTPUT @Agb2200;":ROUT:AGND:CHAN:DIS (@
5-26 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:AGND:PORTInput port is always the input Ground Port. So, the input ports in c
Agilent B2200 User’s Guide, Edition 4 5-27SCPI Command Reference[:ROUTe]:AGND[:STATe]Example OUTPUT @Agb2200;":ROUT:AGND:PORT 0,13"OUTPUT @
5-28 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:AGND:UNUSEDExample OUTPUT @Agb2200;":ROUT:AGND:STAT 0,ON"OUTPUT @A
Agilent B2200 User’s Guide, Edition 4 5-29SCPI Command Reference[:ROUTe]:BIAS:CHANnel:DISable:CARD[:ROUTe]:BIAS:CHANnel:DISable:CARD This command bias
5-30 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:BIAS:CHANnel:ENABle:CARDExample OUTPUT @Agb2200;":ROUT:BIAS:CHAN:ENAB:C
Agilent B2200 User’s Guide, Edition 4 5-31SCPI Command Reference[:ROUTe]:BIAS:PORTInput port is always the input Bias Port. So, the input ports in cha
Agilent B2200 User’s Guide, Edition 4Contents:SYSTem:ERRor? . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-32 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:BIAS[:STATe]In this example, A will be 4.[:ROUTe]:BIAS[:STATe] NOTE You canno
Agilent B2200 User’s Guide, Edition 4 5-33SCPI Command Reference[:ROUTe]:CLOSe:CARD?[:ROUTe]:CLOSe:CARD? This query command returns channel_list of al
5-34 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:CONNection:RULEExample OUTPUT @Agb2200;":ROUT:CLOS (@10101,10202)"
Agilent B2200 User’s Guide, Edition 4 5-35SCPI Command Reference[:ROUTe]:COUPle:PORTThe connection sequence is used only for cards that have been set
5-36 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:COUPle:PORT:DETectAt *RST, no input ports are coupled.Syntax [:ROUTe]:COUPle
Agilent B2200 User’s Guide, Edition 4 5-37SCPI Command Reference[:ROUTe]:COUPle[:STATe][:ROUTe]:COUPle[:STATe] This command controls the Couple Mode f
5-38 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:FUNCtion[:ROUTe]:FUNCtion This command sets the channel configuration, which
Agilent B2200 User’s Guide, Edition 4 5-39SCPI Command Reference[:ROUTe]:OPEN[:LIST][:ROUTe]:OPEN[:LIST]This command disconnects the input ports from
5-40 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference[:ROUTe]:SYMBol:PORTSyntax [:ROUTe]:SYMBol:CHANnel card_number,channel_number,'s
Agilent B2200 User’s Guide, Edition 4 5-41SCPI Command Reference:SYSTem:BEEP[:ROUTe]:SYMBol:PORT? port_numberQuery response symbol_string <newline
ContentsAgilent B2200 User’s Guide, Edition 4agb220xa_cmdData_Q . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-42 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:SYSTem:CDEScription?Syntax :SYSTem:CCONfig? card_numberQuery response card_configur
Agilent B2200 User’s Guide, Edition 4 5-43SCPI Command Reference:SYSTem:CPONFor B2201A:• In the Auto Config mode, "card_description" respons
5-44 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:SYSTem:CTYPe?For the device reset state, see *RST.Syntax :SYSTem:CPON card_numberExa
Agilent B2200 User’s Guide, Edition 4 5-45SCPI Command Reference:SYSTem:DISPlay:LEDSyntax :SYSTem:DISPlay:LCD stateEven though state is OFF, some sta
5-46 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:SYSTem:ERRor?Syntax :SYSTem:DISPlay:STRing 'string'Example OUTPUT @Agb220
Agilent B2200 User’s Guide, Edition 4 5-47SCPI Command Reference:SYSTem:MEMOry:COMMentExample OUTPUT @Agb2200;":SYST:KLC ON":SYSTem:MEMOry:
5-48 Agilent B2200 User’s Guide, Edition 4SCPI Command Reference:SYSTem:MEMOry:LOAD:SYSTem:MEMOry:LOADThis command restores the B2200 setup informatio
Agilent B2200 User’s Guide, Edition 4 5-49SCPI Command Reference:SYSTem:PEN:SYSTem:PEN This command enables/disables the light pen.Syntax :SYSTem:PEN
5-50 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceStatus Reporting StructureStatus Reporting Structure This section describes the status
Agilent B2200 User’s Guide, Edition 4 5-51SCPI Command ReferenceStatus Reporting StructureFigure 5-1 Status Reporting Structure of B2200 The above fig
Agilent B2200 User’s Guide, Edition 4Contentsagb220xa_readStatusByte_Q. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
5-52 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceStatus Byte RegisterStatus Byte Register The B2200 Status Byte Register contains bits
Agilent B2200 User’s Guide, Edition 4 5-53SCPI Command ReferenceStatus Byte RegisterIn general, use serial polling (not *STB?) inside interrupt servic
5-54 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceService Request Enable RegisterService Request Enable Register The Service Request Ena
Agilent B2200 User’s Guide, Edition 4 5-55SCPI Command ReferenceStandard Event Status RegisterStandard Event Status Register The Standard Event Status
5-56 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceStandard Event Status Enable RegisterTable 5-2 Standard Event Status Register of B2200
Agilent B2200 User’s Guide, Edition 4 5-57SCPI Command ReferenceOutput QueueOutput Queue The Output Queue stores response messages until they are read
5-58 Agilent B2200 User’s Guide, Edition 4SCPI Command ReferenceOutput Queue
6 VXIplug&play Driver
6-2 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverThis chapter introduces the VXIplug&play driver available for Agilent B2200, and c
Agilent B2200 User’s Guide, Edition 4 6-3VXIplug&play DriverSystem RequirementsSystem RequirementsThe following system environments are required.•
1 Introduction
6-4 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverInstalling VXIplug&play DriverInstalling VXIplug&play DriverThe installation f
Agilent B2200 User’s Guide, Edition 4 6-5VXIplug&play DriverDriver FunctionsDriver FunctionsTable 6-1 lists all the functions for the Agilent B220
6-6 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsGround Mode agb220xa_groundPort Selects the input ground port of the s
Agilent B2200 User’s Guide, Edition 4 6-7VXIplug&play DriverDriver Functionsagb220xa_biasChanCardThis function bias-enables or disables all the ou
6-8 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_biasChanList_QThis function will query the instrument for the
Agilent B2200 User’s Guide, Edition 4 6-9VXIplug&play DriverDriver Functionsbport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (card
6-10 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_closeCard_QThis function will query the card for the channel
Agilent B2200 User’s Guide, Edition 4 6-11VXIplug&play DriverDriver FunctionsThe parameter “closechan_list” is an array of integers with each inte
6-12 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionscmd_str Instrument command (cannot exceed 256 bytes in length).size
Agilent B2200 User’s Guide, Edition 4 6-13VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_cmdInt16_Q(ViSession vi, ViString c
Notices© Agilent Technologies, Inc. 2004, 2005, 2009, 2011No part of this manual may be reproduced in any form or by any means (including electronic s
1-2 Agilent B2200 User’s Guide, Edition 4IntroductionThis chapter describes the basic functions and features of the Agilent B2200A/B2201A (Agilent B22
6-14 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_cmdRealThis function passes the cmd_str string to the instru
Agilent B2200 User’s Guide, Edition 4 6-15VXIplug&play DriverDriver Functionscmd_str Instrument command (cannot exceed 256 bytes in length).resul
6-16 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_connRuleSeqThe function sets connection rule and connection
Agilent B2200 User’s Guide, Edition 4 6-17VXIplug&play DriverDriver Functionscport_cardno Card number. 1 (card 1), 2 (card 2), 3 (card 3), 4 (car
6-18 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsA device clear will abort the present operation and enable the instru
Agilent B2200 User’s Guide, Edition 4 6-19VXIplug&play DriverDriver FunctionsInstrument errors may occur when you places the instrument in a bad s
6-20 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_esr_Q(ViSession vi, ViChar_VI_FAR e
Agilent B2200 User’s Guide, Edition 4 6-21VXIplug&play DriverDriver Functions1 : sets ground disabled card.gnd_cardno Card number. 1 (card 1), 2
6-22 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_groundChanList_Q(ViSession vi, ViIn
Agilent B2200 User’s Guide, Edition 4 6-23VXIplug&play DriverDriver Functionsgnd_port Input port number to be set to the ground port. 1 to 14 (in
Agilent B2200 User’s Guide, Edition 4 1-3IntroductionAgilent B2200 SeriesAgilent B2200 SeriesAgilent B2200 series is a switching matrix designed for s
6-24 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsvi Instrument handle. This is VI_NULL if an error occurred during th
Agilent B2200 User’s Guide, Edition 4 6-25VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_openList(ViSession vi, ViInt32_VI_F
6-26 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver FunctionsSyntax ViStatus _VI_FUNC agb220xa_readStatusByte_Q(ViSession vi, ViPI
Agilent B2200 User’s Guide, Edition 4 6-27VXIplug&play DriverDriver Functionsfile_name Compensation data file name. Use absolute path. If the val
6-28 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functionsagb220xa_testExec_QThis function performs the diagnostics and returns
Agilent B2200 User’s Guide, Edition 4 6-29VXIplug&play DriverDriver Functionsagb220xa_timeOut_QThis function returns the timeout value for driver
6-30 Agilent B2200 User’s Guide, Edition 4VXIplug&play DriverDriver Functions
7 Error Messages
7-2 Agilent B2200 User’s Guide, Edition 4Error MessagesThis chapter lists and describes the error messages for Agilent B2200. An error message consist
Agilent B2200 User’s Guide, Edition 4 7-3Error MessagesStandard SCPI Error MessagesStandard SCPI Error Messages Standard SCPI error messages have nega
1-4 Agilent B2200 User’s Guide, Edition 4IntroductionFront PanelFront PanelThe Agilent B2200 series provides the front panel keys, the LCD, and the LE
7-4 Agilent B2200 User’s Guide, Edition 4Error MessagesStandard SCPI Error Messages-103 Invalid separator An illegal character was received when a sep
Agilent B2200 User’s Guide, Edition 4 7-5Error MessagesStandard SCPI Error Messages-120 Numeric data error An error was detected in a numeric paramete
7-6 Agilent B2200 User’s Guide, Edition 4Error MessagesStandard SCPI Error Messages-148 Character data not allowed A character parameter is not allowe
Agilent B2200 User’s Guide, Edition 4 7-7Error MessagesStandard SCPI Error MessagesExecution Error If syntax of a SCPI command header and parameter is
7-8 Agilent B2200 User’s Guide, Edition 4Error MessagesStandard SCPI Error MessagesDevice-Dependent Errors -3XX errors indicate that an B2200 operatio
Agilent B2200 User’s Guide, Edition 4 7-9Error MessagesStandard SCPI Error MessagesQuery Errors If the output queue control of the B2200 detects one o
7-10 Agilent B2200 User’s Guide, Edition 4Error MessagesB2200 Specific Error MessagesB2200 Specific Error Messages These are the B2200-specific errors
Agilent B2200 User’s Guide, Edition 4 7-11Error MessagesB2200 Specific Error MessagesB2200 Channel Related Errors Error Number Error Message and Descr
7-12 Agilent B2200 User’s Guide, Edition 4Error MessagesB2200 Specific Error MessagesB2200 Card/Mode/Port Related Errors Error Number Error Message an
Agilent B2200 User’s Guide, Edition 4 7-13Error MessagesB2200 Specific Error Messages3017 Too many relays closed. Max 52 relays/card. Too many relays
Agilent B2200 User’s Guide, Edition 4 1-5IntroductionFront PanelFigure 1-1 Front Panel View
7-14 Agilent B2200 User’s Guide, Edition 4Error MessagesB2200 Specific Error Messages3026 Cannot use same port for Unused and CoupleThe couple mode an
1-6 Agilent B2200 User’s Guide, Edition 4IntroductionRear PanelRear PanelThe Agilent B2200 series has four card slots for the switch modules, the GPIB
Agilent B2200 User’s Guide, Edition 4 1-7IntroductionRear PanelFigure 1-2 Rear Panel View
1-8 Agilent B2200 User’s Guide, Edition 4IntroductionSwitch ModulesSwitch ModulesThe Agilent B2200A and B2201A support the dedicated switch module, Ag
Agilent B2200 User’s Guide, Edition 4 1-9IntroductionSwitch ModulesFigure 1-4 Switch Module Block Diagram OutputInput12345678910111213 CMH14 CML123456
1-10 Agilent B2200 User’s Guide, Edition 4IntroductionSpecificationsSpecificationsThis section lists specifications and supplemental information for t
Agilent B2200 User’s Guide, Edition 4 1-11IntroductionSpecificationsGeneral Specifications Temperature range Operating: 5 °C to 35 °CStorage: -20 °C t
DECLARATION OF CONFORMITY According to EN ISO/IEC 17050-1:2004Manufacturer’s Name: Agilent Technologies Singapore (International) Pte. Ltd. Manufactur
1-12 Agilent B2200 User’s Guide, Edition 4IntroductionSpecificationsSwitch ModulesTable 1-1 lists the specifications when the switch module is install
Agilent B2200 User’s Guide, Edition 4 1-13IntroductionSpecificationsSupplemental Characteristics for B2200A/B2210AOffset current 1 10 fA, I-V portIM
1-14 Agilent B2200 User’s Guide, Edition 4IntroductionSpecificationsSupplemental Characteristics for B2201A/B2211A Offset current 1 100 fA, I-V portI
Agilent B2200 User’s Guide, Edition 4 1-15IntroductionAccessories and OptionsAccessories and OptionsAgilent B2200 is furnished with the following acce
1-16 Agilent B2200 User’s Guide, Edition 4IntroductionAccessories and OptionsTable 1-2 Options and AccessoriesModel NumberOption Item Description B220
Agilent B2200 User’s Guide, Edition 4 1-17IntroductionAccessories and Options16494A Triaxial cable16494A-001 1.5 m length16494A-002 3 m length16494A-0
1-18 Agilent B2200 User’s Guide, Edition 4IntroductionAccessories and Options
2 Installation
2-2 Agilent B2200 User’s Guide, Edition 4InstallationThis chapter describes requirements to install Agilent B2200 and the tasks for installation, and
Agilent B2200 User’s Guide, Edition 4 2-3InstallationRequirementsRequirements This section describes the following requirements for the Agilent B2200.
• HerstellerbescheinigungGERÄUSCHEMISSIONLpa < 70 dBam Arbeitsplatznormaler Betriebnach DIN 45635 T. 19• Manufacturer’s DeclarationACOUSTIC NOISE E
2-4 Agilent B2200 User’s Guide, Edition 4InstallationRequirements• Plug: BS 1363/A, 250 V, 10 A• PN: 8120-4420• Plug: AS 3112, 250 V, 10 A• PN: 8120-4
Agilent B2200 User’s Guide, Edition 4 2-5InstallationRequirementsWARNING For protection from electrical shock, the power cable ground must not be defe
2-6 Agilent B2200 User’s Guide, Edition 4InstallationInspectionInspectionCAUTION Before Opening Packing MaterialsThe Agilent B2200 contains the conden
Agilent B2200 User’s Guide, Edition 4 2-7InstallationInstalling the B2200Installing the B2200 This section describes how to install the Agilent B2200.
2-8 Agilent B2200 User’s Guide, Edition 4InstallationInstalling the B2200To Set the GPIB Address Every device on the GPIB bus must have a unique addre
Agilent B2200 User’s Guide, Edition 4 2-9InstallationInstalling the B2200To Install the Switch ModuleWARNING To prevent electrical shock, turn off the
2-10 Agilent B2200 User’s Guide, Edition 4InstallationSelf-TestSelf-TestNOTE To confirm the specificationsThe self-test and diagnostics checks the ope
Agilent B2200 User’s Guide, Edition 4 2-11InstallationOutput ConnectionsOutput ConnectionsThis section describes how to connect the Agilent B2200 outp
2-12 Agilent B2200 User’s Guide, Edition 4InstallationOutput ConnectionsConnector PlatesConnector plates (Table 2-1) are used for the connection betwe
Agilent B2200 User’s Guide, Edition 4 2-13InstallationOutput ConnectionsFor Kelvin connection, use Kelvin triaxial cable listed in Table 2-2. To make
This product complies with the WEEE Directive (2002/96/EC) marking requirements. The affixed label indicates that you must not discard this electrical
2-14 Agilent B2200 User’s Guide, Edition 4InstallationOutput ConnectionsTable 2-3 Connection to the DUT InterfaceCAUTION Never connect the guard termi
Agilent B2200 User’s Guide, Edition 4 2-15InstallationOutput ConnectionsTo Make Interlock Circuit The interlock circuit is to prevent electric shock w
2-16 Agilent B2200 User’s Guide, Edition 4InstallationOutput ConnectionsFigure 2-4 Dimensions of Interlock Switch (Agilent part number 3101-0302) Figu
Agilent B2200 User’s Guide, Edition 4 2-17InstallationOutput ConnectionsThe 4155/4156 semiconductor parameter analyzer's Intlk connector provides
2-18 Agilent B2200 User’s Guide, Edition 4InstallationOutput ConnectionsNOTE To Check Interlock Circuit If you use the 4155/4156, you can easily check
Agilent B2200 User’s Guide, Edition 4 2-19InstallationOutput ConnectionsTable 2-4 Recommended Parts Table 2-5 Dimensions of Connector Holes Descriptio
2-20 Agilent B2200 User’s Guide, Edition 4InstallationInput ConnectionsInput ConnectionsThis section explains how to connect instruments to the Agilen
Agilent B2200 User’s Guide, Edition 4 2-21InstallationInput ConnectionsTable 2-6 B2200 Input Connections Application Instrument B2200 Input Connector
2-22 Agilent B2200 User’s Guide, Edition 4InstallationInput ConnectionsIn the table above, • CMH/CML connectors are a pair of the AUX inputs.• The Agi
Agilent B2200 User’s Guide, Edition 4 2-23InstallationMeasurement Cable LengthMeasurement Cable Length This section describes how to calculate the tot
Safety SummaryThe following general safety precautions must be observed during all phases of operation, service, and repair of this instrument. Failur
2-24 Agilent B2200 User’s Guide, Edition 4InstallationMeasurement Cable LengthTable 2-7 Guard Capacitances of B2200 Measurement Environment (Typical)
Agilent B2200 User’s Guide, Edition 4 2-25InstallationMaintenanceMaintenanceMaintenance should be performed periodically to keep the B2200 in good con
2-26 Agilent B2200 User’s Guide, Edition 4InstallationMaintenance
3 Front Panel Operation
3-2 Agilent B2200 User’s Guide, Edition 4Front Panel OperationThis chapter explains the front panel operation and the switch control functions of the
Agilent B2200 User’s Guide, Edition 4 3-3Front Panel OperationOperationOperationThis section describes operations of the Agilent B2200.• “To Initializ
3-4 Agilent B2200 User’s Guide, Edition 4Front Panel OperationOperationTo Initialize Agilent B22001. Press the Shift key and the Local key. Then the L
Agilent B2200 User’s Guide, Edition 4 3-5Front Panel OperationOperationTo Change Connection Rule1. Press the Rule key. The LCD displays the present se
3-6 Agilent B2200 User’s Guide, Edition 4Front Panel OperationOperationTo Open All Switches1. Press the Open All key.2. Press the arrow key to display
Agilent B2200 User’s Guide, Edition 4 3-7Front Panel OperationOperationTo Use Bias ModeBias mode cannot be set to ON when the ground mode is ON.1. Cha
Because of the danger of introducing additional hazards, do not install substitute parts or perform any unauthorized modification to the instrument. R
3-8 Agilent B2200 User’s Guide, Edition 4Front Panel OperationOperationTo Use Ground ModeGround mode cannot be set to ON when the bias mode is ON.1. C
Agilent B2200 User’s Guide, Edition 4 3-9Front Panel OperationOperationTo Use Couple Mode1. Press the Couple Mode key to set the couple mode ON. Press
3-10 Agilent B2200 User’s Guide, Edition 4Front Panel OperationOperationTo Read Error Message1. Press the Menu key. The setup menu will be displayed o
Agilent B2200 User’s Guide, Edition 4 3-11Front Panel OperationOperationTo Set Remote Display ModeThis instruction enables or disables the data displa
3-12 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsSwitch Control Functions This section introduces the switch con
Agilent B2200 User’s Guide, Edition 4 3-13Front Panel OperationSwitch Control FunctionsChannel Configuration ModeThe Agilent B2200 provides the two ch
3-14 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsConnection Rule Connection rule is used to specify if an input
Agilent B2200 User’s Guide, Edition 4 3-15Front Panel OperationSwitch Control FunctionsConnection Sequence NOTE This is only for modules that are set
3-16 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsBias Mode Bias mode is useful for connecting the same input to
Agilent B2200 User’s Guide, Edition 4 3-17Front Panel OperationSwitch Control FunctionsNOTE Connection rule cannot be specified for the input bias por
Safety SymbolsThe general definitions of safety symbols used on equipment or in manuals are listed below.Instruction manual symbol: the product will b
3-18 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsGround ModeGround mode operation is similar to the bias mode op
Agilent B2200 User’s Guide, Edition 4 3-19Front Panel OperationSwitch Control FunctionsNOTE Connection rule cannot be specified for the input ground p
3-20 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSwitch Control FunctionsCouple Mode Couple mode is useful for making the Kelvin connect
Agilent B2200 User’s Guide, Edition 4 3-21Front Panel OperationSwitch Control FunctionsNOTE Couple Port Detection FunctionThe Agilent B2200 provides t
3-22 Agilent B2200 User’s Guide, Edition 4Front Panel OperationDisplay FunctionsDisplay FunctionsThe Agilent B2200 provides LED matrix, LCD, and 18 fr
Agilent B2200 User’s Guide, Edition 4 3-23Front Panel OperationDisplay FunctionsLCDThe LCD displays the setup information and the status information a
3-24 Agilent B2200 User’s Guide, Edition 4Front Panel OperationDisplay FunctionsTable 3-5 LCD Display ItemsNOTE In the remote mode and the default set
Agilent B2200 User’s Guide, Edition 4 3-25Front Panel OperationFront Panel KeysFront Panel KeysThe Agilent B2200 provides LED matrix, LCD, and 18 fron
3-26 Agilent B2200 User’s Guide, Edition 4Front Panel OperationFront Panel KeysFunction key groupThe Function key group is used to change the input po
Agilent B2200 User’s Guide, Edition 4 3-27Front Panel OperationFront Panel KeysBias Mode Sets the bias mode ON or OFF. When the bias mode is set to ON
In This Manual This manual is a user’s guide for Agilent B2200A/B2201A Switch Mainframe, and consists of the following chapters:• IntroductionDescribe
3-28 Agilent B2200 User’s Guide, Edition 4Front Panel OperationFront Panel KeysGnd Ch (Shift+Open All)Displays the following menu.SET VIEW• SET displa
Agilent B2200 User’s Guide, Edition 4 3-29Front Panel OperationFront Panel KeysEdit key groupThe Edit key group is used to select the switch module to
3-30 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusSetup MenusThis section explains the following setup menus displayed by pres
Agilent B2200 User’s Guide, Edition 4 3-31Front Panel OperationSetup MenusFunction Tree The setup menu provides the following functions.• CONFIG• ADDR
3-32 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusCONFIG Displays the sub menu that provides the following functions.• ADDRESS
Agilent B2200 User’s Guide, Edition 4 3-33Front Panel OperationSetup MenusERROR Displays the sub menu that provides the following functions.• DISPLAYD
3-34 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusSelftest MenuPress the Shift key and the Menu key. The selftest menu is disp
Agilent B2200 User’s Guide, Edition 4 3-35Front Panel OperationSetup MenusSELF_TEST Displays the sub menu that provides the following functions.• EXEC
3-36 Agilent B2200 User’s Guide, Edition 4Front Panel OperationSetup MenusKEY Displays the sub menu that provides the following functions.• EXECUTEDis
Agilent B2200 User’s Guide, Edition 4 3-37Front Panel OperationSetup MenusLED Displays the sub menu that provides the following functions.• EXECUTESta