15
DC power supply specifications (continued)
Square-wave output characteristics
Table 10. Square-wave output characteristics
Parameter Range Characteristics
Amplitude accuracy
± (offset)
S1 (30 V/1 A) and S1S2 (AUTO) 0.2 V
S2 (8 V/3 A) and S1S2 (AUTO) 0.2 V
Amplitude resolution
S1 (30 V/1 A) 1 mV
S2 (8 V/3 A) 1 mV
Frequency accuracy
± (% of frequency setting + offset)
(27 steps
1
) 0.005% + 0.01 Hz
Frequency resolution 0.01 Hz
Duty cycle accuracy
± (% of duty cycle setting)
(256 steps: 0.39% to 99.60%) 0.4%
2, 3
Duty cycle resolution 0.39%
3
Pulse width accuracy
3, 4
± (offset)
(256 steps: 1/frequency) Duty cycle/frequency
Pulse width resolution Range/256
1. Available frequencies: 0.5, 2, 5, 6, 10, 15, 25, 30, 40, 50, 60, 75, 80, 100, 120, 150, 200, 240, 300, 400, 480, 600, 800, 1200, 1600, 2400, 4800(Hz).
If range S1S2 (AUTO) is selected, available frequencies range is 10 to 4800 Hz, with fixed 50% duty cycle:
Output Range Adjustable step Accuracy
Frequency 10.0 Hz to 4800.0 Hz 10 Hz/100 Hz/1000 Hz
around
0.005% + 0.1 Hz
(according to the display of frequency indication)
2. For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is added. The accuracy of the duty cycle should be calculated as:
Accuracy =
(
0.4% +
[(
frequency
– 1
)
× 0.1%
])
100
Calculation example: Frequency setting = 4800 Hz, Duty cycle setting = 50%
Characteristic of duty cycle = ± 0.4% +
[(
4800
– 1
)
× 0.1%
]
= ±
5.1
= ± 5.1%
100 100
The duty cycle accuracy (for frequency setting 4800 Hz) is calculated as 50% ± 5.1%.
3. Characteristic applies when the positive or negative pulse width is greater than 50 μs.
4. For frequency signals greater than 100 Hz, an additional 0.1% per 100 Hz is added. The accuracy of the pulse width should be calculated as:
(
0.4% +
[(
frequency
– 1
)
× 0.1%
])
100
Accuracy =
frequency
Calculation example: Frequency setting = 4800 Hz, Duty cycle setting = 50%
Characteristic of pulse width = ±
((
0.4% +
[(
4800
– 1
)
× 0.1%
])
×
1
)
= ±
5.1
×
1
)
= ± 10.625 µs
100 4800 100 4800
The pulse width accuracy (for frequency setting 4800 Hz and duty cycle setting 50%) is calculated as 0.1042 ms ±10.625 μs.
The rise and fall time are 25 μs typically between 10% and 90% of the signal amplitude.
The additional load regulation is 0.15 V/A.