Agilent U3606BMultimeter I DC Power Supply Data SheetConvenient and Full Featured. One-box source-and-measure device.Features• Combination of a 5.5 d
10Digital multimeter specifications (continued)Supplemental characteristics (continued)AC currentMeasurement method AC coupled true rmsMaximum input c
11Digital multimeter specifications (continued)Supplemental characteristics (continued)DiodeMeasurement method 0.83 mA ± 0.2% constant current sourceO
12DC power supply specificationsSafety considerations The U3606B is a safety class I instrument, which means it has a protective earth terminal. The t
13DC power supply specifications (continued)Performance specifications (continued)Table 8. DC power supply performance specificationsParameterSpecific
14DC power supply specifications (continued)Supplemental characteristicsTable 9. DC power supply supplemental characteristicsParameterCharacteristicsS
15DC power supply specifications (continued)Square-wave output characteristicsTable 10. Square-wave output characteristicsParameter Range Characterist
16Sweep characteristicsTable 11. Scan output characteristicsScan Constant voltage Constant currentRange S1 S2 S2m S1 S1m S2Maximum amplitude131.500 V
17Product characteristicsPower supply • Universal 100 Vac to 240 Vac ±10%• AC line frequency of 45 to 66 Hz; 360 to 440 Hz for 100/120 V operationPo
18Ordering informationStandard shipped items• Quick Start Guide• Product Reference CD• Agilent IO Library Suite• Certificate of Calibration• U820
Agilent Channel Partnerswww.agilent.com/find/channelpartnersGet the best of both worlds: Agilent’s measurement expertise and product breadth, combined
2Convenient two instruments in one boxLooking for a one-box source-and-measure device to meet your measurement needs? The Agilent U3606B Multimeter |
3Take a closer lookFor DC power supply: Full-featured 30-W DC supply• 30 V/1 A and 8 V/3 A (for U3606A)• 30 V/1 A, 100 mA/30 V, 8V/3 A and 1000 mV/
4Digital multimeter specificationsSpecification assumptions:• Specifications stated are after 60 minutes of warm-up and for 5½-digit resolution• One
5Digital multimeter specifications (continued)DC specifications (continued)Table 1. DC accuracy specifications ± (% of reading + % of range)Function R
6Digital multimeter specifications (continued)AC specificationsTable 2. AC accuracy specifications ± (% of reading + % of range)Function Range1Frequen
7Digital multimeter specifications (continued)Frequency specificationsTable 3. Frequency accuracy specifications ± (% of reading + % of range)Function
8Duty cycle and pulse width specificationsTable 6. Duty cycle and pulse width resolution and accuracyFunction Range Resolution Accuracy of full scaleD
9Digital multimeter specifications (continued)Supplemental characteristicsDC voltageMeasurement method Sigma Delta A-to-D converterMaximum input volta
Komentáře k této Příručce