Additional Measurement Functions
The following MEASure and CONFigure statements can be substituted into
the example programs to make measurements other than DC voltage.
Additional Stand-Alone
Multimeter Functions
The following statements can be substituted into the program “Making a
Single Measurement” on page 42.
!AC voltage.
OUTPUT 70903;"MEAS:VOLT:AC?"
!
4-wire resistance.
OUTPUT 70903;"MEAS:FRES?"
!
4-wire thermistor (type = 2252, 5000, 10000).
OUTPUT 70903;"MEAS:TEMP? FTH,type"
!
4-wire RTD (type = 85, 92).
OUTPUT 70903;"MEAS:TEMP? FRTD,type"
The following statements can be substituted into the programs where the
faceplate terminals are used to make multiple reading bursts or multiple
burst measurements.
!AC voltage.
OUTPUT 70903;"CONF:VOLT:AC"
!
4-wire resistance.
OUTPUT 70903;"CONF:FRES"
!
4-wire thermistor (type = 2252, 5000, 10000).
OUTPUT 70903;"CONF:TEMP FTH,type"
!
4-wire RTD (type = 85, 92).
OUTPUT 70903;"CONF:TEMP FRTD,type"
Additional Scanning
Multimeter Functions
The following statements can be substituted into the program “Scanning a
Channel List” on page 46.
!AC voltage.
OUTPUT 70903;"MEAS:VOLT:AC? (@channel_list)"
!
2-wire resistance.
OUTPUT 70903;"MEAS:RES? (@channel_list)"
!
4-wire resistance (channels available are 00 through 07).
OUTPUT 70903;"MEAS:FRES? (@channel_list)"
!
Thermocouple (type = B, E, J, K, N14, N28, R, S, T).
OUTPUT 70903;"MEAS:TEMP? TC,type,(@channel_list)"
Chapter 3 Using the HP E1326B/E1411B Multimeter 69