Chapter 12 173
Basic Measurements Programming Commands
Basic Transmit Power
Basic Measurements Programming
Commands
Step 4. Press the Single hardkey to run the Sequence.
Basic transmit power measurement SCPI commands
The Basic Transmit Power measurement requires the following parameters to be specified:
•Radio Standard
Digital IF BW
Filter Type
Filter Bandwidth
Filter Alpha
Upper and Lower limits
Radio Standard is specified as part of the Acquisition and is a node in some of the SCPI commands so
does not have an explicit SCPI command. The following commands are used to set up the remaining
parameters:
[:SENSe]:LSEQuencer:BTXPower:[NONE]:DIF:BANDwidth|BWIDth[:RESolution]
<freq>
[:SENSe]:LSEQuencer:BTXPower:[NONE]:DIF:FILTer:TYPE
GAUSsian|FLATtop|RRC|SNYQuist
[:SENSe]:LSEQuencer:BTXPower:[NONE]:DIF:FILTer:BANDwidth|BWIDth
[:RESolution] <freq>
[:SENSe]:LSEQuencer:BTXPower:[NONE]:DIF:FILTer:ALPHa <real>
[:SENSe]:LSEQuencer:BTXPower:LIMit:UPPer:STATe OFF|ON|0|1
[:SENSe]:LSEQuencer:BTXPower:LIMit:UPPer <rel_ampl>
[:SENSe]:LSEQuencer:BTXPower:LIMit:LOWer:STATe OFF|ON|0|1
[:SENSe]:LSEQuencer:BTXPower:LIMit:LOWer <rel_ampl>
For more detailed information about the SCPI commands above refer to the Sequence Analyzer online
help (or in pdf format, the Users and Programmers Reference).