Agilent Technologies InfiniiVision 7000B Series Servisní příručka Strana 63

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 118
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 62
Calibrating and Adjusting 3
7000 Series Oscilloscopes Service Guide 63
BNC(f)-to-BNC(f) (barrel connector) to the tee as shown
below.
Figure 10 User Calibration cable for 4-channel oscilloscope
3 Connect a BNC cable (40 inches maximum) from the TRIG
OUT connector on the rear panel to the BNC barrel
connector.
4 Press the Utility key, then press the Service softkey.
5 Begin the Self Cal by pressing the Start User Cal softkey.
6 When the User Cal is completed, set the rear-panel
CALIBRATION switch to PROTECTED.
Longer cable
to TRIG OUT
To Channel 3
To Channel 2
To Channel 1
To Channel 4
7000 service guide.book Page 63 Thursday, October 16, 2008 12:18 PM
Zobrazit stránku 62
1 2 ... 58 59 60 61 62 63 64 65 66 67 68 ... 117 118

Komentáře k této Příručce

Žádné komentáře