Agilent Technologies Part no. E4418-90059Revision CSeptember 2003Service GuideAgilent TechnologiesE4418B/E4419B Power Meters
Sales and Service Officesx Agilent E4418B/E4419B Service GuideEurope:Agilent TechnologiesTest & MeasurementEuropean Marketing OrganizationP.O. Box
Replaceable PartsAssembly and Disassembly Guidelines5-16 Agilent E4418B/E4419B Service GuideReplacing the Front Panel Power Sensor Cable Assemblies1.
Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-17
Replaceable PartsAssembly and Disassembly Guidelines5-18 Agilent E4418B/E4419B Service GuideReplacing the Front Panel Power Reference Cable Assembly1.
Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-19Replacing the Rear Panel Power Sensor Cable Assemblies (Op
Replaceable PartsAssembly and Disassembly Guidelines5-20 Agilent E4418B/E4419B Service Guide
Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-21Replacing the Rear Panel Power Reference Cable Assembly (O
Replaceable PartsAssembly and Disassembly Guidelines5-22 Agilent E4418B/E4419B Service GuideReplacing the Rear Panel Recorder Output(s) Cable Assembly
Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-23Replacing The Chassis Assembly1. Remove the Power Meter Co
Replaceable PartsAssembly and Disassembly Guidelines5-24 Agilent E4418B/E4419B Service Guide
6Troubleshooting
Sales and Service OfficesAgilent E4418B/E4419B Service Guide xiReturning Your Power Meter for ServiceUse the information in this section if you need t
TroubleshootingIntroduction6-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter enables qualified service personnel to diagnose suspected f
TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-3Figure 6-1: Troubleshooting Guide -Upper Window TTL OutputStartPower Up meterPOSTPa
TroubleshootingIntroduction6-4 Agilent E4418B/E4419B Service GuideASelectPreset/localConfirmMeas/SetupLimitsLimits OnTTL OutputTTL Output On, Fail O/P
TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-5Figure 6-2: Troubleshooting Guide - Lower Window TTL OutputStartPower Up meterPOSTP
TroubleshootingIntroduction6-6 Agilent E4418B/E4419B Service GuideASelectPreset/localConfirmLower Window - usingMeas/SetupLimitsLimits OnTTL OutputTTL
TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-7Figure 6-3: Troubleshooting Guide - TTL InputsStartPower Up meterPOSTPass?Connect S
TroubleshootingIntroduction6-8 Agilent E4418B/E4419B Service GuideABDid powermeter channelszero?Apply TTL logic 1 level to RJ45 connector. pin 5 wrtpi
TroubleshootingIntroductionAgilent E4418B/E4419B Service Guide 6-9BRemove A2 Processor Assy (see chapter 5) to gain accessto the Communications Assemb
TroubleshootingIntroduction6-10 Agilent E4418B/E4419B Service GuideFigure 6-4: Troubleshooting Guide - RS232/422 InterfaceStartPower upmeterPOSTPass?
Measurement Uncertainty Analysis – Instrument Accuracy TestAgilent E4416A/E4417A Service Guide A-1Appendix AMeasurement Uncertainty Analysis – Instrum
About this Guidexii Agilent E4418B/E4419B Service GuideAbout this GuideChapter 1: SpecificationsThis chapter lists the power meter’s specifications an
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-2 Agilent E4416A/E4417A Service GuideMeasurement IntroductionIn t
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-3Uncertainties1. There is no
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-4 Agilent E4416A/E4417A Service GuideThe Measurement Uncertainty
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-5Standard Uncertainty u(xi):T
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-6 Agilent E4416A/E4417A Service GuideWorst case error on 0.1V ran
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-7Worst case error on 15.8V ma
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-8 Agilent E4416A/E4417A Service GuideBy using the DMM Uncertainti
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide A-9Combined Standard Uncertaint
Measurement Uncertainty Analysis – Instrument Accuracy TestMeasurement IntroductionA-10 Agilent E4416A/E4417A Service Guide
Measurement Uncertainty Analysis – Power Reference Level TestAgilent E4416A/E4417A Service Guide B-1Appendix BMeasurement Uncertainty Analysis – Power
Table of ContentsPageAgilent E4418B/E4419B Service Guide Contents-1Specifications...
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-2 Agilent E4416A/E4417A Service GuideMeasurement IntroductionIn
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-3where,Ps is the reflection
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-4 Agilent E4416A/E4417A Service Guide:The equation above includ
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-5A normal distribution is a
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-6 Agilent E4416A/E4417A Service GuideConsider the model below:A
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-7So u(Ms) from the equation
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-8 Agilent E4416A/E4417A Service GuideResistanceThe uncertainty
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-9Table B-1: Measurement Re
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-10 Agilent E4416A/E4417A Service GuideThe readings in Table B-1
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionAgilent E4416A/E4417A Service Guide B-11Using the above values gi
Contents-2 Agilent E4418B/E4419B Service GuideEquipment ... 2-5Test Setup..
Measurement Uncertainty Analysis – Power Reference Level TestMeasurement IntroductionB-12 Agilent E4416A/E4417A Service GuideCombined Standard Uncerta
IndexIndex-1 Agilent E4418B/E4419B Service GuideAA1 power supply/Battery Charger assembly 4-3A2 processor assembly 4-4A3 front panel assembly 4-7A4 in
Index-2 Agilent E4418B/E4419B Service Guidefront panel repair 5-6GGP-IBconnector characteristic 1-14Hhumidity characteristic 1-15IIEC 1010-1 1-viiinst
Agilent E4418B/E4419B Service Guide Index-3replaceable parts 5-2returning for service 1-xiRmt I/O 1-14RS232connector characteristics 1-14RS422connecto
Index-4 Agilent E4418B/E4419B Service Guide
Agilent E4418B/E4419B Service Guide Contents-3Procedure... 3-11Theory of
Contents-4 Agilent E4418B/E4419B Service GuideDefine the Measurement Equation...A-3Sensitivity Coefficients Ci:...
1Specifications
SpecificationsIntroduction1-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter details the power meter’s specifications and supplemental ch
SpecificationsPower Meter SpecificationsAgilent E4418B/E4419B Service Guide 1-3Power Meter SpecificationsMeterFrequency Range100 kHz to 110 GHz, power
ii Agilent E4418B/E4419B Service Guide© Copyright 1998, 2003Agilent TechnologiesAll rights reserved. Reproduction, adaptation, or translation without
SpecificationsPower Meter Specifications1-4 Agilent E4418B/E4419B Service GuideAccuracyInstrumentationAbsolute: ±0.02 dB (Logarithmic) or ±0.5% (Linea
SpecificationsPower Meter SpecificationsAgilent E4418B/E4419B Service Guide 1-51 mW Power Reference1• Power Output: 1.00 mW (0.0 dBm). • Accuracy: (fo
SpecificationsPower Meter Supplemental Characteristics1-6 Agilent E4418B/E4419B Service GuidePower Meter Supplemental CharacteristicsMeasurement Speed
SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-7Measurement NoisePower sensor dependent (refer to Table 1
SpecificationsPower Meter Supplemental Characteristics1-8 Agilent E4418B/E4419B Service GuideTable 1-3: Power Sensor Specifications±Power Sensor Zero
SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-9Settling Time0 to 99% settled readings over the GP-IB.For
SpecificationsPower Meter Supplemental Characteristics1-10 Agilent E4418B/E4419B Service GuideFor Agilent E-series power sensorsIn FAST mode, within t
SpecificationsPower Meter Supplemental CharacteristicsAgilent E4418B/E4419B Service Guide 1-11Auto filter, default resolution, 10 dB decreasing power
SpecificationsPower Meter Supplemental Characteristics1-12 Agilent E4418B/E4419B Service GuidePower Sensor SpecificationsDefinitionsZero SetIn any pow
SpecificationsBattery Option 001 Operational CharacteristicsAgilent E4418B/E4419B Service Guide 1-13Battery Option 001 Operational CharacteristicsThe
Legal InformationAgilent E4418B/E4419B Service Guide iiiLegal InformationNoticeInformation contained in this document is subject to change without not
SpecificationsGeneral Characteristics1-14 Agilent E4418B/E4419B Service GuideGeneral CharacteristicsRear Panel ConnectorsRecorder Output(s)Analog 0-1
SpecificationsEnvironmental CharacteristicsAgilent E4418B/E4419B Service Guide 1-15Environmental CharacteristicsGeneral ConditionsComplies with the re
SpecificationsGeneral1-16 Agilent E4418B/E4419B Service GuideGeneralDimensionsThe following dimensions exclude front and rear panel protrusions:212.6
SpecificationsGeneralAgilent E4418B/E4419B Service Guide 1-17Remote ProgrammingInterfaceGP-IB interface operates to IEEE 488.2. RS-232 and RS-422 inte
SpecificationsGeneral1-18 Agilent E4418B/E4419B Service Guide
2Performance Tests
Performance TestsIntroduction2-2 Agilent E4418B/E4419B Service GuideIntroductionThe procedures in this chapter test the power meter’s electrical perfo
Performance TestsIntroductionAgilent E4418B/E4419B Service Guide 2-3Equipment RequiredTable 2-1 lists all the equipment required for the adjustments a
Performance TestsIntroduction2-4 Agilent E4418B/E4419B Service GuidePerformance Test RecordResults of the performance tests may be tabulated in Table
Performance TestsZero TestAgilent E4418B/E4419B Service Guide 2-5Zero Test SpecificationDescriptionAfter the power meter is initially calibrated and z
Legal Informationiv Agilent E4418B/E4419B Service GuideLimitation of Warranty The foregoing warranty shall not apply to defects resulting from imprope
Performance TestsZero Test2-6 Agilent E4418B/E4419B Service GuideProcedureThe following procedure should be performed for the Zero Test. Note The pr
Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-7Instrument Accuracy Test SpecificationDescriptionThe power meter accur
Performance TestsInstrument Accuracy Test2-8 Agilent E4418B/E4419B Service GuideTest SetupFigure 2-2: Instrument Accuracy Test SetupProcedureThe foll
Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-9Note When switching the range calibrator to STANDBY, allow enough ti
Performance TestsInstrument Accuracy Test2-10 Agilent E4418B/E4419B Service GuideTable 2-3: Instrument Accuracy Results (For Indication Only)Note T
Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-11Test Procedure 2Note This test procedure makes use of a 11683A Rang
Performance TestsInstrument Accuracy Test2-12 Agilent E4418B/E4419B Service GuideProcedureThe following procedure should be performed for the Instrume
Performance TestsInstrument Accuracy TestAgilent E4418B/E4419B Service Guide 2-13Note When switching the range calibrator to STANDBY, allow enough t
Performance TestsInstrument Accuracy Test2-14 Agilent E4418B/E4419B Service Guide14. Set the DC calibrator output to the values shown in Table 2-4. Fo
Performance TestsPower Reference Level TestAgilent E4418B/E4419B Service Guide 2-15Power Reference Level Test DescriptionThis test verifies the power
Equipment OperationAgilent E4418B/E4419B Service Guide vEquipment OperationWarnings and CautionsThis guide uses warnings and cautions to denote hazard
Performance TestsPower Reference Level Test2-16 Agilent E4418B/E4419B Service GuideIf an equivalent measurement system is employed for post-factory te
Performance TestsPower Reference Level TestAgilent E4418B/E4419B Service Guide 2-17Equipment• Test power meter: Agilent 432A.• Thermistor mount: Agile
Performance TestsPower Reference Level Test2-18 Agilent E4418B/E4419B Service Guide4. Preset the power meter:Press , then on the power meter.Set th
Performance TestsAgilent E-Series Power Sensor Interface TestAgilent E4418B/E4419B Service Guide 2-19Agilent E-Series Power Sensor Interface TestThe A
Performance TestsAgilent E-Series Power Sensor Functional Test2-20 Agilent E4418B/E4419B Service GuideAgilent E-Series Power Sensor Functional TestDes
Performance TestsAgilent E-Series Power Sensor Functional TestAgilent E4418B/E4419B Service Guide 2-212. Press then .3. On the power meter press ,
Performance TestsAgilent E-Series Power Sensor Functional Test2-22 Agilent E4418B/E4419B Service GuideTable 2-6: Functional Test ResultTestChannel A
Performance TestsPerformance Test RecordAgilent E4418B/E4419B Service Guide 2-23Performance Test RecordModel Agilent E4418B/E4419B Power MeterTested b
Performance TestsPerformance Test Record2-24 Agilent E4418B/E4419B Service GuideOutput Standing Wave Ratio Test1N/A___________ ___________1.061, 21.08
3Adjustments
Equipment Operationvi Agilent E4418B/E4419B Service GuideGeneral Safety ConsiderationsWARNING Before this instrument is switched on, make sure it ha
AdjustmentsIntroduction3-2 Agilent E4418B/E4419B Service GuideIntroduction This chapter describes adjustments and checks which ensure proper performan
AdjustmentsIntroductionAgilent E4418B/E4419B Service Guide 3-3Equipment RequiredThe adjustment procedures include a list of recommended test equipment
AdjustmentsPower Reference Oscillator Frequency Adjustment3-4 Agilent E4418B/E4419B Service GuidePower Reference Oscillator Frequency AdjustmentNote
AdjustmentsPower Reference Oscillator Frequency AdjustmentAgilent E4418B/E4419B Service Guide 3-53. Turn the power reference on by pressing , on th
AdjustmentsPower Reference Oscillator Level Adjustment3-6 Agilent E4418B/E4419B Service GuidePower Reference Oscillator Level AdjustmentDescriptionThi
AdjustmentsPower Reference Oscillator Level AdjustmentAgilent E4418B/E4419B Service Guide 3-7Note The power meter may be returned to the nearest Agi
AdjustmentsPower Reference Oscillator Level Adjustment3-8 Agilent E4418B/E4419B Service GuideBy substituting equation 3 into equation 1 and manipulati
AdjustmentsPower Reference Oscillator Level AdjustmentAgilent E4418B/E4419B Service Guide 3-93. Connect the equipment as shown in Figure 3-3. The lead
AdjustmentsPower Reference Oscillator Level Adjustment3-10 Agilent E4418B/E4419B Service Guide14.Remove the power meter’s cover and adjust A2R90 until
AdjustmentsDisplay Brightness and Contrast AdjustmentAgilent E4418B/E4419B Service Guide 3-11Display Brightness and Contrast AdjustmentIntroductionThe
Equipment OperationAgilent E4418B/E4419B Service Guide viiIEC 1010-1 ComplianceThis instrument has been designed and tested in accordance with IEC Pub
AdjustmentsDisplay Brightness and Contrast Adjustment3-12 Agilent E4418B/E4419B Service Guide
4Theory of Operation
Theory of OperationIntroduction4-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter describes how each of the power meter’s assemblies oper
Theory of OperationA1 Power Supply/Battery ChargerAgilent E4418B/E4419B Service Guide 4-3A1 Power Supply/Battery ChargerThe A1 power supply/battery ch
Theory of OperationA2 Processor Assembly4-4 Agilent E4418B/E4419B Service GuideA2 Processor AssemblyThe processor assembly contains the microcontrolle
Theory of OperationA2 Processor AssemblyAgilent E4418B/E4419B Service Guide 4-5The LCD controller on the A3 front panel assembly is configured as a me
Theory of OperationA2 Processor Assembly4-6 Agilent E4418B/E4419B Service GuideThe microcontroller circuits that control all the above functions, and
Theory of OperationA3 Front Panel AssemblyAgilent E4418B/E4419B Service Guide 4-7A3 Front Panel AssemblyThe front panel assembly is made up of a liqui
Theory of OperationA4 Comms Assembly4-8 Agilent E4418B/E4419B Service GuideA4 Comms AssemblyThe comms assembly contains the circuitry required for rem
Theory of OperationA5 Daughter AssemblyAgilent E4418B/E4419B Service Guide 4-9A5 Daughter AssemblyThe A5 daughter assembly is loaded vertically into t
List of Related Publicationsviii Agilent E4418B/E4419B Service GuideList of Related PublicationsThe Agilent E4418B and Agilent E4419B User’s Guides ar
Theory of OperationA6 Measurement Assembly4-10 Agilent E4418B/E4419B Service GuideA6 Measurement AssemblyThere is one measurement assembly in the Agil
Theory of OperationA6 Measurement AssemblyAgilent E4418B/E4419B Service Guide 4-11The DSP chip controls the logic which sets the chopper driver voltag
Theory of OperationA8 Rechargeable Battery Assembly4-12 Agilent E4418B/E4419B Service GuideA8 Rechargeable Battery Assembly The A8 Rechargeable Batter
Theory of OperationA8 Rechargeable Battery AssemblyAgilent E4418B/E4419B Service Guide 4-13Remove this page and replace with pullout.fm.
Theory of OperationA8 Rechargeable Battery Assembly4-14 Agilent E4418B/E4419B Service Guide
5Replaceable Parts
Replaceable PartsIntroduction5-2 Agilent E4418B/E4419B Service GuideIntroductionThis chapter contains details of some of the higher level components a
Replaceable PartsAssembly Part NumbersAgilent E4418B/E4419B Service Guide 5-3Assembly Part NumbersReference DesignatorDescriptionAgilent Part NumberA1
Replaceable PartsAssembly Part Numbers5-4 Agilent E4418B/E4419B Service GuideA304 Front Panel Assembly for:Agilent E4419B Option 003New E4419-61002Ref
Replaceable PartsAssembly Part NumbersAgilent E4418B/E4419B Service Guide 5-5Power Supply Assembly TypesBefore ordering an A1 Power Supply Assembly, c
Sales and Service OfficesAgilent E4418B/E4419B Service Guide ixSales and Service OfficesFor more information about Agilent Technologies test and measu
Replaceable PartsFront Panel Connector Repair5-6 Agilent E4418B/E4419B Service GuideFront Panel Connector RepairThe front panel assembly is an exchang
Replaceable PartsFirmware UpgradesAgilent E4418B/E4419B Service Guide 5-7Firmware UpgradesThe power meter’s firmware can be upgraded using a PC. The c
Replaceable PartsAssembly and Disassembly Guidelines5-8 Agilent E4418B/E4419B Service GuideAssembly and Disassembly GuidelinesThe guidelines in this s
Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-9Removing the A1 Power Supply Assembly1. Remove the power su
Replaceable PartsAssembly and Disassembly Guidelines5-10 Agilent E4418B/E4419B Service GuideRemoving the A2 Processor Assembly1. Remove the A5 daughte
Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-11Removing the A4 Comms Assembly1. Remove the A2 processor,
Replaceable PartsAssembly and Disassembly Guidelines5-12 Agilent E4418B/E4419B Service GuideRemoving the A5 Daughter or A6 Measurement Assemblies1. Di
Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-13Removing the A7 Fan Assembly1. Remove the A1 power supply
Replaceable PartsAssembly and Disassembly Guidelines5-14 Agilent E4418B/E4419B Service GuideRemoving the Power Meter RAM Battery (BT1)1. Remove the A2
Replaceable PartsAssembly and Disassembly GuidelinesAgilent E4418B/E4419B Service Guide 5-15Removing the A3 Front Panel Assembly1. Disconnect the foll