Agilent TechnologiesEPM Series Power MetersE-Series and 8480 Series Power SensorsTechnical SpecificationsProduct Specifications and CharacteristicsE44
10The chart in Figure 1 shows the typicaluncertainty in making a relative powermeasurement, using the same powermeter channel and the same powersensor
E-series E9300 Average Power Sensor SpecificationsThe E-series E9300 wide dynamic range, average power sensors are designed for use with the EPM famil
12Typical SWR, 10 MHz to 18 GHz, (25 ˚C ± 10 ˚C) for E9300A and E9301A sensorsTypical SWR, 9kHz to 6 GHz, (25 ˚C ± 10 ˚C) for E9304A sensors1.181.21.1
13Typical SWR, 10 MHz to 18 GHz, (25 ˚C ± 10 ˚C) for E9300B and E9301B sensorsTypical SWR, 10 MHz to 18 GHz, (25 ˚C ± 10 ˚C) for E9300H and E9301H sen
14Power Linearity (after zero and cal at ambient environmental conditions)Table 10Sensor Power Linearity Linearity(25 ˚C ± 10 ˚C) (0 – 55 ˚C)E9300A, E
15E-Series E9300 Average Power SensorsTypical E9300B/01B Power Linearity at 25 ˚C, after zero and calibration, with associatedMeasurement UncertaintyP
16Figure 2 shows the typical uncertainty in making a relative power measurement, using the same power meter channel andsame power sensor to obtain the
17E-Series E9300 Average Power Sensors Switch Point Data:The E9300 power sensors have two paths, a low power path covers: -60 to -10 dBm ("A"
18Calibration Factor (CF) and Reflection Coefficient (Rho)Calibration Factor and Reflection Coefficient data are provided at frequency intervals on a
19The 8480 series power sensors aredesigned for use with the 435B, 436A,437B, 438A, 70100A, E1416A and nowthe E4418B and E4419B power meters.Thes
2Table 1AccuracyInstrumentationAbsolute: ±0.02 dB (Log) or ±0.5%(linear). Please add the correspondingpower sensor linearity percentagefrom Tables 7 a
Model Frequency Range Maximum SWR Power Linearity (1)Maximum Power Connector Type Weight8485A 50 MHz to 26.5 GHz 50 MHz to 100 MHz: 1.15 +10 dBm to +
21High Sensitivity Sensors, 100 pW to 10 µW (–70 dBm to –20 dBm)1. Negligible deviation except for those power ranges noted. 2. For pulses greater th
For more information about Agilent Technologies test and measurement products, applications, services,and for a current sales office listing, visit ou
3Supplemental CharacteristicsPower ReferenceFrequency: 50 MHz nominal.SWR: 1.05 maximum.Connector: Type N (f), 50 ohms.Measurement SpeedThree measurem
4Auto filter, default resolution, 10 dB decreasing power step (Normal and x2 modes):For E-Series Sensors:In FAST mode (using Free Run trigger), withi
X2 NormalMode ModeMaximum Sensor Power40 ms 70 msTypicalSensorSettlingDynamicTimesRange-40 dBm70 ms 120 ms-50 dBm210 ms 400 ms-60 dBm3.4 s 6.5 sX2 Nor
6Power Meter FunctionsAccessed by key entry: either hardkeys, or softkey menu, and programmable.Zero: Zeros the meter. (Power refer-ence calibrator is
71. Characteristics describe product performance that is useful in the application of the product, but is not covered by the product warranty.Battery
8Options AvailableOption 001: Supplies internal rechargeable battery.Option 002: Supplies parallel rearpanel sensor input(s). (Power reference calibra
Power LinearityTable 7Power Temperature (25 ºC ±5 C) Temperature (0 to 55 ºC)100 pW to 10 mW ±3 % ±7 %(-70 dBm to +10 dBm)10 mW to 100 mW ±4.5 % ±10 %