 Agilent 1000 Series OscilloscopesUser’s Guide

Strany 2

10 Agilent 1000 Series Oscilloscopes User’s Guide Contents5 Saving, Recalling, and Printing Data 113Saving and Recalling Data 114To save and recall wa

Strany 3

4 Making Measurements100 Agilent 1000 Series Oscilloscopes User’s GuideVoltage MeasurementsThere are 10 automatic voltage measurements:• Vmax (Maximu

Strany 4 - In This Book

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 101Vmin (Minimum Voltage)The minimum amplitude. The most negative peak voltage me

Strany 5

4 Making Measurements102 Agilent 1000 Series Oscilloscopes User’s GuideVrms (Root-Mean-Square Voltage)The true root- mean- square voltage over the en

Strany 6

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 103Time MeasurementsThere are 12 automatic time measurements plus the hardware fr

Strany 7

4 Making Measurements104 Agilent 1000 Series Oscilloscopes User’s GuideFrequencyMeasures the frequency of a waveform. See Figure 41 on page 103.Rise

Strany 8

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 105Positive Pulse WidthMeasures the positive pulse width of a waveform.Negative P

Strany 9

4 Making Measurements106 Agilent 1000 Series Oscilloscopes User’s GuideDelay Between Rising EdgesMeasures the delay between two waveforms using the r

Strany 10 - Contents

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 107Phase Between Rising EdgesMeasures the phase between two waveforms using the r

Strany 11

4 Making Measurements108 Agilent 1000 Series Oscilloscopes User’s GuideCounter (Frequency)The 1000 Series oscilloscopes have an integrated 6- digit h

Strany 12 - Index 153

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 109Making Cursor MeasurementsYou can use the [Cursors] front panel key to select

Strany 13

ContentsAgilent 1000 Series Oscilloscopes User’s Guide 11 Performing Mask Tests 131To enable/disable mask tests 131To select the source channel for ma

Strany 14

4 Making Measurements110 Agilent 1000 Series Oscilloscopes User’s GuideTo use manually adjustable cursorsYou can set up two parallel, manually adjust

Strany 15

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 111To use tracking cross-hair cursorsYou can set up one or two manually adjustabl

Strany 16

4 Making Measurements112 Agilent 1000 Series Oscilloscopes User’s GuideTo display cursors for automatic measurements1 Press [Cursors].2 In the Curso

Strany 17 - Getting Started

113Agilent 1000 Series OscilloscopesUser’s Guide 5Saving, Recalling, and Printing DataSaving and Recalling Data 114Using the Disk Manager 118Printi

Strany 18

5 Saving, Recalling, and Printing Data114 Agilent 1000 Series Oscilloscopes User’s GuideSaving and Recalling DataUsing the oscilloscope’s [Save/Recal

Strany 19

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 1153 Continue pressing the Storage softkey or turn the entry k

Strany 20

5 Saving, Recalling, and Printing Data116 Agilent 1000 Series Oscilloscopes User’s GuideTo save to or recall from internal storage:a Press Internal.

Strany 21 - Step 4. Input a waveform

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 1174 To specify whether oscilloscope parameters be saved along w

Strany 22 - Step 5. Use Auto-Scale

5 Saving, Recalling, and Printing Data118 Agilent 1000 Series Oscilloscopes User’s GuideUsing the Disk ManagerWhen a USB drive is connected to the fr

Strany 23 - Getting Started 1

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 119To switch between files, path, and directory panes1 In the Di

Strany 24 - Step 6. Compensate probes

12 Agilent 1000 Series Oscilloscopes User’s Guide ContentsIndex 153

Strany 25 - High Frequency Compensation

5 Saving, Recalling, and Printing Data120 Agilent 1000 Series Oscilloscopes User’s GuideTo edit folder/file namesIn the folder/file name edit dialog:

Strany 26 - DSO1024A

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 121To delete foldersIn the directory pane (see “To switch betwee

Strany 27

5 Saving, Recalling, and Printing Data122 Agilent 1000 Series Oscilloscopes User’s GuideTo recall filesIn the files pane (see “To switch between file

Strany 28

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 123Printing ScreensYou can print oscilloscope display screens to

Strany 29

5 Saving, Recalling, and Printing Data124 Agilent 1000 Series Oscilloscopes User’s GuideTo choose a PictBridge printerYou can print to a PictBridge c

Strany 30 - Figure 8 Oscilloscope Display

Saving, Recalling, and Printing Data 5Agilent 1000 Series Oscilloscopes User’s Guide 125To print with inverted screen colors1 Press [Print].2 In the

Strany 31

5 Saving, Recalling, and Printing Data126 Agilent 1000 Series Oscilloscopes User’s GuideTo copy a screen to the printer1 Press [Print].2 In the Prin

Strany 32

127Agilent 1000 Series OscilloscopesUser’s Guide 6Oscilloscope Utility SettingsDisplaying System Information 128Turning Sound ON or OFF 128Setting

Strany 33 - Hole for Kensington lock

6 Oscilloscope Utility Settings128 Agilent 1000 Series Oscilloscopes User’s GuideDisplaying System InformationTo display the oscilloscope’s system in

Strany 34

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 129Setting and Displaying the Date and TimeTo set and display the oscil

Strany 35 - Displaying Data

Agilent 1000 Series Oscilloscopes User’s Guide 13 FiguresFigure 1. Power Switch 19Figure 2. [Default Setup] Key 20Figure 3. [Auto-Scale] Key 22Figure

Strany 36 - Position knob

6 Oscilloscope Utility Settings130 Agilent 1000 Series Oscilloscopes User’s GuideSetting the Language (Menu and Help)To set the language used in menu

Strany 37

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 131Performing Mask TestsThe mask test function monitors waveform change

Strany 38 - Slow Scan Mode

6 Oscilloscope Utility Settings132 Agilent 1000 Series Oscilloscopes User’s GuideTo run/stop a mask test1 In the Mask Test menu ([Utility]>Mask

Strany 39

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 133To set the mask test output condition1 In the Mask Test menu ([Utili

Strany 40 - Zoomed window view

6 Oscilloscope Utility Settings134 Agilent 1000 Series Oscilloscopes User’s GuideTo adjust a mask’s horizontal failure margin1 In the Mask menu ([Uti

Strany 41 - X-Y Format

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 135To select internal/external mask storage location1 In the Mask menu

Strany 42 - To view the sample rate

6 Oscilloscope Utility Settings136 Agilent 1000 Series Oscilloscopes User’s GuideSetting PreferencesThe oscilloscope’s Preference menu lets you set s

Strany 43 - Math REF

Oscilloscope Utility Settings 6Agilent 1000 Series Oscilloscopes User’s Guide 137To select the USB device port functionThe (square) USB device port o

Strany 44 - To adjust the vertical scale

6 Oscilloscope Utility Settings138 Agilent 1000 Series Oscilloscopes User’s GuideRunning Self-CalibrationThe automatic calibration routine adjusts th

Strany 45 - To specify channel coupling

139Agilent 1000 Series OscilloscopesUser’s Guide 7Specifications and CharacteristicsEnvironmental Conditions 140Measurement Category 141Specificati

Strany 46 - AC Coupling Status

14 Agilent 1000 Series Oscilloscopes User’s Guide FiguresFigure 28. Sample Rate and Oscilloscope Bandwidth 69Figure 29. Real-Time Sampling Mode 71Figu

Strany 47 - Bandwidth limit OFF

7 Specifications and Characteristics140 Agilent 1000 Series Oscilloscopes User’s GuideEnvironmental ConditionsOvervoltage CategoryThis product is int

Strany 48 - Bandwidth ON Status

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 141Measurement CategoryThe 1000 Series oscilloscope is intended to

Strany 49 - To use a digital filter

7 Specifications and Characteristics142 Agilent 1000 Series Oscilloscopes User’s GuideSpecificationsNOTEAll specifications are warranted. Specificati

Strany 50 - To invert a waveform

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 143CharacteristicsNOTEAll characteristics are the typical performa

Strany 51 - Displaying Data 2

7 Specifications and Characteristics144 Agilent 1000 Series Oscilloscopes User’s GuideAC coupled1: DSO102xA: 5 Hz to 200 MHzDSO101xA: 5 Hz to 100 MHz

Strany 52 - To specify the channel units

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 145Table 8 Horizontal System CharacteristicsName Typical ValueRang

Strany 53 - Math Scale

7 Specifications and Characteristics146 Agilent 1000 Series Oscilloscopes User’s GuideTable 9 Trigger System CharacteristicsName Typical ValueSources

Strany 54

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 147Table 11 Measurement FeaturesName Typical ValueAutomatic measur

Strany 55

7 Specifications and Characteristics148 Agilent 1000 Series Oscilloscopes User’s GuideTable 13 StorageName Typical ValueSave/Recall internal: 10 setu

Strany 56 - FFT Resolution

Specifications and Characteristics 7Agilent 1000 Series Oscilloscopes User’s Guide 149Table 15 General CharacteristicsName Typical ValuePhysical size

Strany 57 - To save a reference waveform

Agilent 1000 Series Oscilloscopes User’s Guide 15 TablesTable 1. Agilent 1000 Series Oscilloscope Models 3Table 2. Auto-Scale Default Settings 23Table

Strany 58

7 Specifications and Characteristics150 Agilent 1000 Series Oscilloscopes User’s GuideThe N2739A rack mount kit is available for placing a 1000 Serie

Strany 59 - Changing the Display Settings

151Agilent 1000 Series OscilloscopesUser’s Guide ASafety NoticesWarnings 151Safety Symbols 152This apparatus has been designed and tested in

Strany 60 - To set waveform persistence

ASafety Notices 152 Agilent 1000 Series Oscilloscopes User’s Guide• Whenever it is likely that the ground protection is impaired, you must make the

Strany 61 - To change the grid

Agilent 1000 Series Oscilloscopes User’s Guide 153 IndexNumerics50% trigger level, 83AAC channel coupling, 45AC coupling, 54AC trigger coupling, 87, 9

Strany 62 - To invert screen colors

154 Agilent 1000 Series Oscilloscopes User’s GuideIndex Fall Time measurement, 104FFT (Fast Fourier Transform) math function, 53, 54FFT resolution, 56

Strany 63 - To select screen persistence

Agilent 1000 Series Oscilloscopes User’s Guide 155Index nonvolatile memory locations, 113Normal acquisition mode, 73normal adjustment, 44NTSC standard

Strany 64 - 2 Displaying Data

156 Agilent 1000 Series Oscilloscopes User’s GuideIndex system information, displaying, 128Ttemperature at last calibration, 142theory, sampling, 66ti

Strany 66 - Aliasing

Agilent Technologies, Inc. 2008-2009Printed in Malaysia 10/09Second Edition, October 2009*54130-97014*54130-97014www.agilent.com

Strany 67

16 Agilent 1000 Series Oscilloscopes User’s Guide Tables

Strany 68

17Agilent 1000 Series OscilloscopesUser’s Guide 1Getting StartedStep 1. Inspect the package contents 18Step 2. Turn on the oscilloscope 19Step 3. L

Strany 69 - Oscilloscope Rise Time

1Getting Started18 Agilent 1000 Series Oscilloscopes User’s GuideStep 1. Inspect the package contents1 Inspect the shipping container for damage.Keep

Strany 70

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 19Step 2. Turn on the oscilloscopeThe next few steps (turning on the oscilloscope, lo

Strany 71 - Real-Time Sampling

Agilent 1000 Series Oscilloscopes User’s GuideNotices© Agilent Technologies, Inc. 2008-2009No part of this manual may be reproduced in any form or by

Strany 72 - Memory Depth and Sample Rate

1Getting Started20 Agilent 1000 Series Oscilloscopes User’s GuideStep 3. Load the default oscilloscope setupYou can recall the factory default setup

Strany 73 - Choosing the Acquisition Mode

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 21Step 4. Input a waveform1 Input a waveform to a channel of the oscilloscope.Use one

Strany 74

1Getting Started22 Agilent 1000 Series Oscilloscopes User’s GuideStep 5. Use Auto-ScaleThe oscilloscope has an auto- scale feature that automatically

Strany 75

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 23The oscilloscope is configured to the following default control settings:Table 2 Au

Strany 76

1Getting Started24 Agilent 1000 Series Oscilloscopes User’s GuideStep 6. Compensate probesCompensate probes to match your probe to the input channel.

Strany 77

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 25High Frequency CompensationFor the supplied passive probes:1 Using the BNC adapter,

Strany 78 - To record waveforms

1Getting Started26 Agilent 1000 Series Oscilloscopes User’s GuideStep 7. Become familiar with the Front Panel ControlsBefore using the oscilloscope,

Strany 79 - To play-back waveforms

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 27The definitions of the front panel knobs, keys, and softkeys are as follows:Front P

Strany 80 - To select the end frame

1Getting Started28 Agilent 1000 Series Oscilloscopes User’s GuideUsing the Oscilloscope Softkey MenusWhen one of the oscilloscope front panel keys tu

Strany 81 - To store recorded waveforms

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 29The [Menu On/Off] front panel key turns off the menu or turns on the last accessed

Strany 82 - To import/export recordings

Agilent 1000 Series Oscilloscopes User’s Guide 3 Agilent 1000 Series Oscilloscopes—At a GlanceThe Agilent 1000 Series oscilloscopes are low-cost porta

Strany 83 - Trigger level knob

1Getting Started30 Agilent 1000 Series Oscilloscopes User’s GuideStep 8. Become familiar with the oscilloscope displayChannel 1 Status,DC coupling, 5

Strany 84 - To force a trigger

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 31Step 9. Use the Run Control keysThere are two front panel keys for starting and sto

Strany 85 - To set up edge triggers

1Getting Started32 Agilent 1000 Series Oscilloscopes User’s GuideStep 10. Access the built-in helpThe oscilloscope has built- in quick help informati

Strany 86

Getting Started 1Agilent 1000 Series Oscilloscopes User’s Guide 33Securing the OscilloscopeTo secure a 1000 Series oscilloscope to its location, you

Strany 87 - To set up video triggers

1Getting Started34 Agilent 1000 Series Oscilloscopes User’s Guide

Strany 88

35Agilent 1000 Series OscilloscopesUser’s Guide 2Displaying DataUsing the Horizontal Controls 36Using the Vertical Controls 43Using Math Function W

Strany 89 - Capturing Data 3

2 Displaying Data36 Agilent 1000 Series Oscilloscopes User’s GuideUsing the Horizontal ControlsThe horizontal controls consist of:• The horizontal sc

Strany 90 - To set up pattern triggers

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 37To adjust the horizontal scale• Turn the horizontal scale knob to change the horizo

Strany 91 - To set up alternate triggers

2 Displaying Data38 Agilent 1000 Series Oscilloscopes User’s GuideWhen the time/div is set to 50 ms/div or slower, the oscilloscope enters Slow Scan

Strany 92 - To set the trigger coupling

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 39• Push the horizontal position knob to “zero” the trigger point (in other words, mo

Strany 93

4 Agilent 1000 Series Oscilloscopes User’s Guide In This BookThis guide shows how to use the Agilent 1000 Series oscilloscopes.1Getting StartedDescrib

Strany 94

2 Displaying Data40 Agilent 1000 Series Oscilloscopes User’s GuideTo change the horizontal time base (Y-T, X-Y, or Roll)1 Press [Menu/Zoom].2 In the

Strany 95 - To specify a trigger holdoff

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 41X-Y FormatThis format compares the voltage level of two waveforms point by point. I

Strany 96 - 3Capturing Data

2 Displaying Data42 Agilent 1000 Series Oscilloscopes User’s GuideThe following modes or functions are not available in X-Y format:• Automatic voltag

Strany 97 - Making Measurements

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 43Using the Vertical ControlsThe vertical controls consist of:• The channel ([1], [2

Strany 98

2 Displaying Data44 Agilent 1000 Series Oscilloscopes User’s GuideTo adjust the vertical scaleWhen an input channel waveform is on:• Turn its vertica

Strany 99

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 45Notice that, as you adjust the vertical position, a message showing the position of

Strany 100 - Vmax (Maximum Voltage)

2 Displaying Data46 Agilent 1000 Series Oscilloscopes User’s GuideFigure 17 DC Coupling ControlFigure 18 AC Coupling ControlDC CouplingDC Coupling St

Strany 101

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 47To specify a bandwidth limitWhen high frequency components of a waveform are not im

Strany 102 - Preshoot

2 Displaying Data48 Agilent 1000 Series Oscilloscopes User’s GuideTo specify the probe attenuationFor correct measurements, you must match the oscill

Strany 103 - Time Measurements

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 49To use a digital filterYou can apply a digital filter to the sampled waveform data.

Strany 104 - Rise Time

Agilent 1000 Series Oscilloscopes User’s Guide 5 ContentsAgilent 1000 Series Oscilloscopes—At a Glance 3In This Book 4Figures 13Tab les 151 Getting St

Strany 105 - Negative Duty Cycle

2 Displaying Data50 Agilent 1000 Series Oscilloscopes User’s GuideTo change the Volts/Div control sensitivityWhen you need to adjust the amplitude/di

Strany 106 - Delay Between Falling Edges

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 51Figure 21 Waveform Before InversionFigure 22 Waveform After Inversion

Strany 107 - Phase Between Falling Edges

2 Displaying Data52 Agilent 1000 Series Oscilloscopes User’s GuideTo specify the channel units1 If the channel’s menu is not currently displayed, pre

Strany 108 - Counter (Frequency)

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 53Using Math Function WaveformsThe math functions control allows the selection of the

Strany 109 - Making Cursor Measurements

2 Displaying Data54 Agilent 1000 Series Oscilloscopes User’s GuideTo add, subtract, or multiply waveforms1 Press [Math].2 In the Math menu, press Op

Strany 110

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 555 Press Window, and continue pressing the softkey or turn the entry knob to sele

Strany 111

2 Displaying Data56 Agilent 1000 Series Oscilloscopes User’s Guide10 Use the horizontal position knob to adjust the frequency per division.The freque

Strany 112

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 57Using Reference WaveformsYou can save a reference waveform to an internal, nonvolat

Strany 113

2 Displaying Data58 Agilent 1000 Series Oscilloscopes User’s Guide3 Presh Location to choose “External”.4 Press Save or Import.5 Use the disk manag

Strany 114 - Saving and Recalling Data

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 59Changing the Display SettingsTo display waveforms as vectors or dots1 Press [Displ

Strany 115

6 Agilent 1000 Series Oscilloscopes User’s Guide ContentsSecuring the Oscilloscope 332 Displaying Data 35Using the Horizontal Controls 36To adjust the

Strany 116

2 Displaying Data60 Agilent 1000 Series Oscilloscopes User’s GuideTo clear the display1 Press [Display].2 In the Display menu, press Clear.To set wa

Strany 117

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 61To display graded waveform intensityWhile the oscilloscope is running, waveforms sh

Strany 118 - Using the Disk Manager

2 Displaying Data62 Agilent 1000 Series Oscilloscopes User’s GuideTo change the menu display timeThe menu display time is how long menus remain on th

Strany 119 - To create new folders

Displaying Data 2Agilent 1000 Series Oscilloscopes User’s Guide 63To select screen persistenceThe screen persistence setting specifies what is displa

Strany 120 - To edit folder/file names

2 Displaying Data64 Agilent 1000 Series Oscilloscopes User’s Guide

Strany 121 - To delete files

65Agilent 1000 Series OscilloscopesUser’s Guide 3Capturing DataOverview of Sampling 66Choosing the Acquisition Mode 73Choosing the Acquisition Mode

Strany 122 - To display disk information

3Capturing Data66 Agilent 1000 Series Oscilloscopes User’s GuideOverview of SamplingTo understand the oscilloscope’s sampling and acquisition modes,

Strany 123 - USB Host Port

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 67Oscilloscope Bandwidth and Sample RateAn oscilloscope’s bandwidth is typically descr

Strany 124

3Capturing Data68 Agilent 1000 Series Oscilloscopes User’s GuideHowever, digital signals have frequency components above the fundamental frequency (s

Strany 125

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 69So, in practice, an oscilloscope’s sample rate should be four or more times its band

Strany 126 - 1 Press [Print]

ContentsAgilent 1000 Series Oscilloscopes User’s Guide 7 Changing the Display Settings 59To display waveforms as vectors or dots 59To clear the displa

Strany 127 - Oscilloscope Utility Settings

3Capturing Data70 Agilent 1000 Series Oscilloscopes User’s GuideAn oscilloscope’s rise time is not the fastest edge speed that the oscilloscope can a

Strany 128 - Turning Sound ON or OFF

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 71Real-Time SamplingThe 1000 Series oscilloscopes provide real- time sampling. In othe

Strany 129

3Capturing Data72 Agilent 1000 Series Oscilloscopes User’s GuideMemory Depth and Sample RateThe number of points of oscilloscope memory is fixed (exc

Strany 130

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 73Choosing the Acquisition ModeThe osilloscope can operate in normal, average, or peak

Strany 131 - To enable/disable mask tests

3Capturing Data74 Agilent 1000 Series Oscilloscopes User’s GuideTo select the Average acquisition modeIn the Average acquisition mode, acquisitions a

Strany 132 - To run/stop a mask test

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 75The Average acquisition mode decreases the screen refresh rate.To select the Average

Strany 133 - To set up masks

3Capturing Data76 Agilent 1000 Series Oscilloscopes User’s GuideHowever, in the Peak Detect acquisition mode, acquisitions are made at the fastest sa

Strany 134

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 77To turn OFF/ON sine(x)/x interpolationWhen sample points are displayed as vectors (i

Strany 135 - To export/import masks

3Capturing Data78 Agilent 1000 Series Oscilloscopes User’s GuideRecording/Playing-back WaveformsYou can record waveforms from input channels or from

Strany 136 - To set up the screen saver

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 79To start/stop recording1 In the Sequence menu ([Acquire]>Sequence>Mode=Rec

Strany 137

8 Agilent 1000 Series Oscilloscopes User’s Guide ContentsAdjusting the Trigger Level 83To adjust the trigger level 83To force a trigger 84Choosing the

Strany 138 - Running Self-Calibration

3Capturing Data80 Agilent 1000 Series Oscilloscopes User’s GuideTo select continuous or one-time play-back1 In the Sequence menu ([Acquire]>Sequ

Strany 139

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 81To store recorded waveformsTo store recorded waveforms:1 Press [Acquire].2 In the A

Strany 140 - Pollution Degree Definitions

3Capturing Data82 Agilent 1000 Series Oscilloscopes User’s GuideTo load a recording1 In the Sequence menu ([Acquire]>Sequence>Mode=Storage)

Strany 141 - Measurement Category

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 83Adjusting the Trigger LevelTo adjust the trigger level• Turn the trigger [Level] kno

Strany 142 - Specifications

3Capturing Data84 Agilent 1000 Series Oscilloscopes User’s GuideTo force a triggerTo make an acquisition even if no valid trigger has been found:1 Pr

Strany 143 - Characteristics

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 85Choosing the Trigger ModeThe trigger determines when captured data should be stored

Strany 144

3Capturing Data86 Agilent 1000 Series Oscilloscopes User’s Guide5 Press Source and continue pressing the softkey or turn the entry knob to select

Strany 145

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 875 Press Source and continue pressing the softkey or turn the entry knob to select

Strany 146

3Capturing Data88 Agilent 1000 Series Oscilloscopes User’s Guide5 Press Polarity to toggle between:6 Press Sync and continue pressing the softkey or

Strany 147 - Table 11 Measurement Features

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 89Figure 35 Line SynchronizationFigure 36 Field Synchronization

Strany 148 - Table 14 I/O

ContentsAgilent 1000 Series Oscilloscopes User’s Guide 9 Voltage Measurements 100Vmax (Maximum Voltage) 100Vmin (Minimum Voltage) 101Vpp (Peak-to-Peak

Strany 149 - Table 16 Power Requirements

3Capturing Data90 Agilent 1000 Series Oscilloscopes User’s GuideTo set up pattern triggers1 Press [Menu].2 In the Trigger menu, press Mode.3 Continu

Strany 150

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 91To set up alternate triggersThe Alternate trigger mode splits the display horizontal

Strany 151 - Warnings

3Capturing Data92 Agilent 1000 Series Oscilloscopes User’s GuideSetting Other Trigger ParametersThese are trigger system parameters that apply in all

Strany 152 - Safety Symbols

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 934 Continue pressing the the Coupling softkey or turn the entry knob to select one

Strany 153

3Capturing Data94 Agilent 1000 Series Oscilloscopes User’s GuideTo change the trigger sensitivityTrigger sensitivity specifies the vertical change th

Strany 154

Capturing Data 3Agilent 1000 Series Oscilloscopes User’s Guide 95To specify a trigger holdoffTrigger holdoff can be used to stabilize a waveform. The

Strany 155

3Capturing Data96 Agilent 1000 Series Oscilloscopes User’s GuideUsing the External Trigger InputYou can trigger on external inputs by selecting “EXT”

Strany 156

97Agilent 1000 Series OscilloscopesUser’s Guide 4Making MeasurementsDisplaying Automatic Measurements 98Voltage Measurements 100Time Measurements 1

Strany 157

4 Making Measurements98 Agilent 1000 Series Oscilloscopes User’s GuideDisplaying Automatic MeasurementsYou can use the [Measure] key to display autom

Strany 158 - *54130-97014*

Making Measurements 4Agilent 1000 Series Oscilloscopes User’s Guide 99A maximum of three measurements can be displayed at the bottom of the display.