Localizing the Problem
Localizing the Problem - Flow Chart (Power-Up)
Localizing the Problem - Flow Chart (Power-Up)
See "Power Up Checks", in this Chapter, for details of the steps given in the flow chart
below.
Figure 1-1 Localizing the Problem - Flow Chart
Power On
Fan On?
Messages
OK?
Beep after
Keys &
Controls OK?
6 seconds?
NO
YES
NO
NO
NO
NO
YES
YES
YES
YES
Goto;
"Troubleshooting the
Power Supply"
Goto;
"Power-Up Self Test
Diagnostics"
Goto;
"If Power-Up
Happened Correctly"
Failure
Reported by
Diagnostics ?
Goto the relevant trouble shooting section;
"Trouble Shooting The Controller/Display" - Chapter 5.
"Trouble Shooting The Power Supply" - Chapter 6.
"Running Diagnostics" - Chapter 2