Agilent Technologies 8922S Specifikace Strana 88

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 244
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 87
8-18
Assembly and Disassembly Procedures
A23 Input Section Removal
A23 Input Section Removal
Done with instrument top and bottom cover removed.
1. Do steps 1 through 11 of the A1 Front Panel removal instructions.
NOTE The front panel assembly must be separated from the main chassis. Considerable pulling
force is required to pull the front panel from the chassis.
2. Remove two 5/8-inch hex nuts.
3. Remove two TX-15 side mounting screws.
4. Remove one TX-15 bottom mounting screw.
5. Disconnect all cabling and remove input section assembly.
Tools Required
TX-15 screw driver
TX-10 screw driver
2-pt. Pozidriv
5/8-inch wrench
1/4-inch wrench
4
1/4" SMC
CONNECTOR
2
(2 places)
3
FRONT PANEL
AND SIDE VIEW
A22 Display
RIBBON
CABLE
1/4" SMC CONNECTOR
BOTTOM VIEW
Zobrazit stránku 87
1 2 ... 83 84 85 86 87 88 89 90 91 92 93 ... 243 244

Komentáře k této Příručce

Žádné komentáře