Agilent PXT Wireless Communications Test Set
User’s Guide
230
Functional Check
This functional check is only available if you have Option N6051A RF Parametric Test
with Test Mode Signaling installed on your PXT.
Turn on the instrument. Wait until the boot procedure completes: indicated by the appearance of the Base
Station Emulator startup screen. With no cabling connected to the front-panel, perform the following:
Press the Amp key and set the RF1 output power to 0 dBm.
Press the FREQ key and set the RF1 DL frequency to match the UL frequency.
Press the SPECTRUM key. The display should be similar to the picture below. This indicates
that the LTE transmit and receive functions are operational.
Figure 7-1: Functional Test Verification