Agilent Technologies E5500 Series Uživatelská příručka Strana 111

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 232
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 110
AC Modes 6
Agilent 5500 SPM User’s Guide 111
MAC III offers additional lock-in amplifiers for other more complex
modes as well). Specially coated MAC cantilevers are also required.
Standard MAC Mode
In standard MAC Mode, the coil is located in the sample plate
(Figure 76). A Contact Mode or AC Mode nose assembly can be used.
Figure 76 MAC Mode sample plate
The procedure for imaging in MAC Mode is the same as for AC Mode,
with these exceptions:
1 Connect the 6-pin (MAC) end of the EC/MAC Cable (Figure 77) to
the 6-pin connector on the sample plate. Connect the other end of the
cable to the EC/MAC socket on the bottom of the microscope stand.
Figure 77 EC/MAC cable connections
Zobrazit stránku 110
1 2 ... 106 107 108 109 110 111 112 113 114 115 116 ... 231 232

Komentáře k této Příručce

Žádné komentáře