Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Adjusting Intermodule Skew
Placing Markers for an Interval Reading.
1. From the Waveform display, select the Markers tab.
2. Place one Time marker on the leading edge of the scope glitch.
3. Place a second Time marker on the leading edge of the analyzer glitch.
4. Read the difference between the two Time markers.
Adjust the Skew
Set the skew setting of the oscilloscope to the time difference you
measured between the two Time marker, then select Apply.
The result of the skew adjustment moves the oscilloscope signal over
to the right (positive) so that it visually aligns itself with the analyzer