Chapter 1: Agilent Technologies 16700A/B-Series Logic Analysis System
Overview - Multiple Frames Configuration
Three frames; Three analyzers; Group Run and
Group Run OR Trigger
In this Multi-frame example, Analyzer 1 triggers and passes the arm
signal to Multi-frame 2 (Mf2) and Mf3. Mf2 then passes the arm signal
to to Analyzer 2. Mf3 then passes the arm signal to the High Speed
Timing Analyzer (HST).
In addition to the Group Run Arming Tree configuration shown below,
configure analyzers 1 and 2 with the OR Trigger macro.
Three frames; Three analyzers; Group Run OR
Trigger
In this Multi-frame example, Analyzer 2 triggers and passes the arm
signal to Multi-frame 2 (Mf2). Mf2 then passes the arm signal to Mf1
and Mf3. Mf1 then passes the arm signal to Analyzer 1. Mf3 then passes
the arm signal to the High Speed Timing Analyzer (HST).