Agilent E6432A Microwave Synthesizer Users Guide 4-187
HPE6432_SetIfAtten
ViStatus HPE6432_SetIfAtten (ViSession instrumentHandle,
ViUInt16 ifAttenuation);
Purpose
This function sets the IF upconverter attenuator and has a range of 0 to
30 dB in 2 dB steps.
Parameter List
instrumentHandle
Variable Type ViSession
This is the ViSession handle that is obtained from the Initialize
function. The instrumentHandle identifies a particular instrument
session.
ifAttenuation
Variable Type ViUInt16
This parameter specifies the amount of IF Attenuation to be applied.
Allowable values: 0 to 30 dB in 2 dB steps
Return Value
This parameter reports the return status of this function.
To obtain further information about the status that is returned, call the
Error Message function or refer to “Error-Code and Fail-Code
Messages” on page 3-93.