Agilent E6432A Microwave Synthesizer Users Guide 4-209
HPE6432_SetQGainDac
ViStatus HPE6432_SetQGainDac (ViSession instrumentHandle,
ViUInt16 qGainDac);
Purpose
This function sets a value for the Q Gain DAC.
Q Gain is used to enter compensation values for internal gain
impairments for the Q signal path. Since it is the ratio of I Gain to
Q Gain that is important, it is usually only necessary to make
adjustments to I Gain or Q Gain, but not both.
Factory Preset Value: 2048 (Recommended)
Parameter List
instrumentHandle
Variable Type ViSession
This is the ViSession handle that is obtained from the Initialize
function. The instrumentHandle identifies a particular instrument
session.
qGainDac
Variable Type ViUInt16
This parameter specifies a value for the Q Gain DAC.
Allowable values: 0 to 4095
Return Value
This parameter reports the return status of this function.
To obtain further information about the status that is returned, call the
Error Message function or refer to “Error-Code and Fail-Code
Messages” on page 3-93.
Related Topics
HPE6432_GetQGainDac