$JLOHQW0$996HULHV6HUYLFH*XLGH
5HSDLULQJWKH$JLOHQW0RGHO9&79&7
Repairing the Agilent
Model V24CT/V26CT
5HPRYLQJWKH6\VWHP%RDUG
Remove the Front Housing. See 9-2.
Disconnect the cables from the System Board as shown below.
Using a small (1/8 inch) flat-blade screwdriver, pry the service port cover from the Rear
Housing and remove.
1
2
3
4
5
6
7
8
9
10
11
3
System Board interface
connectors:
1. Keypad connector
2. Flat Panel Data Connector
3. 5V and 12V power
supply connector
4. Service ports (not cabled)
5. SDN interface cable
6. Defib sync connector
(not cabled)
7. Front-end link connector
8. Speaker connector
9. 60 VDC power supply
connector
10. RS232 connector (with
option J13)
11. Docking Connector Cable
Zobrazit stránku 237