Agilent Technologies DisplayPort Servisní příručka Strana 217

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 246
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 216
Calibrating the Infiniium Oscilloscope and Probe 24
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 217
Figure 82 Oscilloscope Calibration Window
5 Follow the on- screen instructions:
a You will be prompted to disconnect everything from all the inputs:
click the OK button.
b Then you will be prompted to connect the calibration cable with
SMA adapters between the Aux Out and a specified input. Install the
SMA adapter by pressing it on input BNC, and hand tightening the
outer ring turning right. Click the OK button after connecting the
cable as prompted.
c During the calibration of channel 1, you will be prompted a
Calibration Options screen, as shown in
Figure 83 below.
Zobrazit stránku 216
1 2 ... 212 213 214 215 216 217 218 219 220 221 222 ... 245 246

Komentáře k této Příručce

Žádné komentáře