Agilent Technologies DisplayPort Servisní příručka Strana 223

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 246
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 222
Calibrating the Infiniium Oscilloscope and Probe 24
Agilent U7232A DisplayPort Electrical Performance Compliance Test Application 223
c Follow the on- screen instructions.
d At the end of the Atten/Offset Calibration perform the Skew
Calibration.
Differential Probe Head Skew Calibration
This procedure ensures that the timing skew errors between channels are
minimized. Perform the following steps:
1 Referring to Figure 89 below, perform the following steps:
a Select the Start Skew Calibration button and follow the on- screen
instructions. For more information on proper connection of probe to
the oscilloscope, refer to the De- skew and Calibration manual. This
Figure 88 Probe Calibration Window.
Zobrazit stránku 222
1 2 ... 218 219 220 221 222 223 224 225 226 227 228 ... 245 246

Komentáře k této Příručce

Žádné komentáře