Agilent Technologies J-BERT N4903B Servisní příručka Strana 152

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 406
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 151
This command is applicable for M8061A. The command sets the clock source to
either CLK or AUX.
The query retunrs the present state.
OUTPut[1]:MULTiplexer:DATA:XOVer[?]
OUTPut[1]:MULTiplexer:DATA:XOVer <NRf>
OUTPut[1]:MULTiplexer:DATA:XOVer?
This command is applicable for M8061A. The command sets the eye crossover of
the pattern generator's Data Out port.
The query returns the current crossover setting.
OUTPut[1]:MULTiplexer:INTerference:CMODe[:STATe][?]
OUTPut[1]:MULTiplexer:INTerference:CMODe ON | OFF
OUTPut[1]:MULTiplexer:INTerference:CMODe?
This command is applicable for M8061A. The command enables/disables the
common mode interference (CMI) feature. The common mode interference
(asymmetrical mode) relates to ground.
The query returns the present state.
OUTPut[1]:MULTiplexer:INTerference:CMODe:GAIN[?]
OUTPut[1]:MULTiplexer:INTerference:CMODe:GAIN <NRf>
OUTPut[1]:MULTiplexer:INTerference:CMODe:GAIN?
This command is applicable for M8061A. The command sets the gain of the
common mode interference.
The query returns the present setting.
OUTPut[1]:MULTiplexer:INTerference:DMODe[:STATe][?]
OUTPut[1]:MULTiplexer:INTerference:DMODe ON | OFF
5 SCPI Command Reference
152 Agilent J-BERT N4903B High-Performance Serial BERT
Description
Syntax
Description
Syntax
Description
Syntax
Description
Syntax
Zobrazit stránku 151