M3000A/M3046AM3015A/M3016A Service Guide
Testing the Instrument 4-3
Recommendations for Test Frequency
The testing checklist appears in the next section of this chapter. The listed procedures should
be performed as indicated in the Suggested Testing Timetable below. The checklist may be
photocopied and should be completed by the tester. It should be filed for future reference.
6XJJHVWHG7HVWLQJ7LPHWDEOH )UHTXHQF\
Functionality Assurance
Performance Assurance Test
System Check
System Self-Test
Daily prior to using the
Instrument to Monitor a
patient.
Preventive Maintenance Tests
NBP Calibration
Sidestream CO
2
Calibration
Once a year.
Once a year or after
4,000 hours continuous
use and following any
Instrument repairs or the
replacement of any
Instrument parts.
Performance and Safety Tests
Temperature Accuracy
ECG/Resp Performance
Invasive Pressure Perform-
ance
SpO
2
Performance
Mainstream CO
2
Performance
Nurse Call Relay Performance
ECG Sync Performance
Once every two years.
including Safety Checks (in
accordance with IEC 601-1)
System Enclosure Leakage
Current
Protective Earth
Patient Leakage Current
Once every two years.