M3000A/M3046A/M3015A/M3016A Service Guide
Introduction to the Instrument 1-49
Functional Description of the M3016A Measurement Server
Extension Hardware
The Extension receives information signals (such as Temp/Press and mainstream CO
2
) from
the patient then transmits the data through the Server to the Monitor via the Server-to-
Monitor link bar. See the Functional Description of the Measurement Server Hardware earlier
in this chapter for a description of the normal processing of measurement data.
The mainstream CO
2
Extension is always used with a Server. To function correctly, both the
Server and the Extension must have compatible software revisions.
:$51,1*
1HYHUXVHD0HDVXUHPHQW([WHQVLRQZLWKDPHDVXUHPHQWVHUYHUZKLFKFRQWDLQVUHOHDVH
$VRIWZDUH$[[[[6LQFHWKHVRIWZDUHUHTXLUHGWRSURFHVVGDWDIURPWKH([WHQVLRQLV
DEVHQWSDWLHQWVDIHW\FRXOGQRWEHVDIHJXDUGHG
7RYLHZWKHVRIWZDUHUHYLVLRQSUHVVWKH6HWXSNH\DQGVHOHFW5HYLVLRQV
Pairing the Server with the Extension allows for the following:
Measurement of both temperature and invasive blood pressure for the same patient.
Two temperature measurements for the same patient including a Delta Temp calculation
feature.
Two invasive blood pressure measurements for the same patient.
For more information on the Temperature and Invasive Blood Pressure (Temp/Press)
measurement see the section earlier in this chapter.