Agilent Technologies AGILENT U2761A Servisní příručka Strana 53

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 169
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 52
Output Configuration Commands 4
U2761A Programmer’s Reference 45
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
Syntax
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]
<amplitude> [VPP|VRMS|DBM]
This command sets the output amplitude with optional unit.
[SOURce:]VOLTage[:LEVel][:IMMediate][:AMPLitude]?
This query returns the output amplitude for the currently selected
function and unit. Values are always returned in the units set by the most
recent [SOURce:]VOLTage[:LEVel][:IMMediate]:UNIT command.
Parameter
Remarks
The RMS and dBm values can be computed as follows:
Vrms = Vpeak / Crest factor
dBm = 10 × log
10
(P / 0.001), where P = Vrms
2
/ R
Load
[1]
Item Type Range of values Default value
amplitude Numeric 40 mVpp to 5 Vpp (Into 50 Ω load)
80 mVpp to 10 Vpp (Into open circuit)
1 Vpp
[1] For the load resistance values, refer to “OUTPut:LOAD” on page 54.
SOUR : VOLT space
?
amplitude:LEV :IMM : AMPL VPP
VRMS
DBM
Zobrazit stránku 52
1 2 ... 48 49 50 51 52 53 54 55 56 57 58 ... 168 169

Komentáře k této Příručce

Žádné komentáře