Chapter 3: Testing Performance
To Test the Time Interval Accuracy
2 Set up the Format tab.
a In the Analyzer setup window, select the Format tab.
b Under the Format tab, select Pod Assignment.
c In the Pod Assignment window, use the mouse to drag Pods 1 and 2 to
the Analyzer 1 column. Use the mouse to drag pods 3 and 4 to the
Unassigned column.
d Select Close to close the Pod Assignment window.
e Under the Format tab, select the field showing the channel assignments
for Pod 1. Using the mouse, first clear the channels (all “.”), then select
channel 0. An asterisk means that the channel is turned on.
f Select the Threshold field under Pod1. Click on the checkbox next
Apply Threshold Settings to all pods to deselect.