Agilent Technologies 24A Uživatelská příručka Strana 148

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 453
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 147
2-17
Making Mixer Measurements (Option 089 Only)
Conversion Loss Using the Frequency Offset Mode
Performing a Power Meter Calibration Over the RF Range
1. Make the connections as shown in Figure 2-14 for the one-sweep power meter
calibration over the RF range.
Figure 2-14 Power Meter Calibration Equipment Setup for the RF Range
2. Use the power meter settings that were set up in steps 1 through 4 of the “Performing a
Power Meter (Source) Calibration Over the IF Range” section.
3. Perform a one-sweep power meter calibration over the RF frequency range at 0 dBm by
pressing:
( , 8722ES)
NOTE Do not reduce the number of points to perform this power meter calibration.
Reducing the number of points will turn off the receiver calibration.
Cal
PWRMTR CAL
ONE SWEEP
0 x1 10 x1
TAKE CAL SWEEP
Zobrazit stránku 147
1 2 ... 143 144 145 146 147 148 149 150 151 152 153 ... 452 453

Komentáře k této Příručce

Žádné komentáře