6-28
Calibrating for Increased Measurement Accuracy
One-Port Reflection Error Correction
13.To compute the error coefficients, press:
The analyzer displays the corrected data trace. The analyzer also shows the notation
Cor to the left of the screen, indicating that the correction is switched on for this
channel.
NOTE The open, short, and load could be measured in any order, and need not follow
the order in this example.
NOTE You can save or store the error correction to use for later measurements.
Refer to Chapter 4 , “Printing, Plotting, and Saving Measurement Results”
for procedures.
14.This completes the one-port correction for reflection measurements. You can connect
and measure your device under test.
DONE: 1-PORT CAL