7-46
Operating Concepts
Measurement Calibration
Device Measurement
Now the unknown is measured to obtain a value for the measured response, S
11M
, at each
frequency. Refer to Figure 7-33.
Figure 7-33 Measured S
11
This is the one-port error model equation solved for S
11A
. Since the three errors and S
11M
are now known for each test frequency, S
11A
can be computed as follows:
For reflection measurements on two-port devices, the same technique can be applied, but
the test device output port must be terminated in the system characteristic impedance.
This termination should have as low a reflection coefficient as the load used to determine
directivity. The additional reflection error caused by an improper termination at the test
device's output port is not incorporated into the one-port error model.
Two-Port Error Model (ES Models Only)
The error model for measurement of the transmission coefficients (magnitude and phase)
of a two-port device is derived in a similar manner. The potential sources of error are
frequency response (tracking), source match, load match, and isolation as shown in Figure
7-34. These errors are effectively removed using the full two-port error model.
S
11A
S
11M
E
DF
()
E
SF
S
11M
E
DF
()
E
RF
+
------------------------------------------------------------------=