System Board
Assembly
Threshold
Adjustment
1. Disconnect power cord from HP 1650B/51B.
2. Connect negative (-) lead of voltmeter to TP GND. Refer to figure 4-3 for
testpoint and adjustment locations.
3. Connect positive (+ ) lead of voltmeter to HP 1650B A1TP3, or HP 1651B
A1TP2 on the System Board Assembly.
4. Connect power cord to logic analyzer and turn instrument power to ON.
5. Assign HP 1650B POD 3, or HP 1651B POD 2, to a machine in the System
Configuration menu by using front-panel knob and SELECT key.
6. Set User defined Pod Threshold of the pod assigned in the previous step to
-9.9 V in the Format Specification menu by using the following steps if
unfamiliar with the menus.
Figure 4-3. System Board Testpoints and Adjustments
HP 1650B/1651B Adjustments
Service Manual 4-7