11 Press the Tolerance softkey, then turn the knob closest to the
Cursors
key to set the tolerance to ±0.0%.
If additional tolerance is desired, set the tolerance to the appropriate level.
This will be the amount "added on" to the previously created envelope.
12
Press the Create Mask softkey to create the mask from the autostore
information.
Automask Using Autostore Operating Hint
The Automask function takes all the information displayed in half bright to create
the mask. However, the display grid and the autostore information shares the
same half-bright display. If the grid is turned on, and Autostore information is on
the screen when the Create Mask softkey is pressed, a warning message is
displayed: "Grid must be None to generate mask with Autostore." The Display
Grid must be turned to None prior to creating the autostore data in order to use
Automask function. Turning the grid to None after the autostore data is created
erases both the grid and the autostore data. Use of cursors does not affect the
Automask function and is highly recommended to ensure the proper testing
margin in the autostore information.
If there is noise rising on the limit lines, you can use the smooth function in the
mask editor to smooth out the noise.
Operating the Measurement/Storage Module
To create a mask template using Autostore
2–32