Setting Up Generic and Advanced Signals Setting Up a Repetitive Burst Signal
52 Agilent 81133A/81134A Pulse Generator User’s Guide, September 2008
Setting Up a Repetitive Burst
Signal
Task Set up a signal with 4 repeated bursts.
Each burst consists of a 32- bit pattern signal with NRZ data
output format at 100 MHz to check a digital circuit with LVPECL
logic.
The data pattern is:
11100000000000000000000000000000
Using the Graphical User Interface
Instrument Settings To set the instrument parameters:
1 Disable the outputs.
2 In the instrument panel, choose the RBurst mode.
3 Enter 4 for the number of null data packets.
4 Define that the number of zeros that follows the repeated bursts
is as large as the burst length.
5 Set the frequency to 100 MHz.
Data Data Data Data Data Data Data
Data Length = 32 x 10 ns
Burst Length = 4
x Data Length
Number of Null data packets =
Burst Length