Chapter 1 11
Calibrating
Calibration Cycle
Calibration Cycle
The performance tests in chapter 2 should be used to check the analyzer
against its specifications once every year. Specifications are listed in the
Specifications Guide.
Performance Verification Test Tables
The tables on the following pages list the performance tests in Chapter
2 , “Performance Verification Tests, required for each model number.
Perform all the tests marked with a dot in the “Std” (standard) column.
If any options are installed in the analyzer, also perform all tests
marked with a dot in the appropriate option column.