Agilent Technologies E1401-90493 Uživatelský manuál Strana 349

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 642
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 348
Chapter 2 349
Performance Verification Tests
43. Tracking Generator Level Flatness: Agilent E4401B and E4411B (Option 1DN or 1DQ)
performance verification test record.
10.Locate the most positive Level Flatness reading in Table 2-95 and
Table 2-96 for frequencies
1 MHz and 10 MHz and enter this value
as Test Record entry 3 of the performance verification test record.
11.Locate the most negative Level Flatness reading in Table 2-95 and
Table 2-96 for frequencies
1 MHz and 10 MHz and enter this value
as Test Record entry 4 of the performance verification test record.
12.Locate the most positive Level Flatness reading in Table 2-95 for
frequencies
10 MHz and 1.5 GHz and enter this value as Test
Record entry 5 of the performance verification test record.
13.Locate the most negative Level Flatness reading in Table 2-95 for
frequencies
10 MHz and 1.5 GHz and enter this value as Test
Record entry 6 of the performance verification test record.
Zobrazit stránku 348
1 2 ... 344 345 346 347 348 349 350 351 352 353 354 ... 641 642

Komentáře k této Příručce

Žádné komentáře