280 Chapter 2
Performance Verification Tests
32. Spurious Responses: Agilent E4404B, E4405B, E4407B, and E4408B
Figure 2-50 Second Harmonic Distortion Test Setup
3. Preset the synthesized sweeper and set the controls as follows:
CW, 300 MHz
POWER LEVEL, 0dBm
4. On the analyzer, press Peak Search (or Search), Marker, Delta.
5. Record the power meter reading at 300 MHz in Table 2-76.
6. Set the synthesized sweeper CW to 600 MHz.
7. On the analyzer, press
FREQUENCY, 600 MHz, then Peak Search (or
Search).
8. Adjust the synthesized sweeper power level until the marker delta (
Mkr1) amplitude reads 0 dB
±0.10 dB.
9. Enter the power sensor 600 MHz calibration factor into the power
meter.
10.Record the power meter reading at 600 MHz in Table 2-76.
Zobrazit stránku 279