Module 6
Low Current Measurement
6-6
z Sweep 0 to 1 volt with no cables attached to SMU
z Check that there is a +/-3 fA base line
z Add prober cables but no prober
z Wait several minutes
z Still +/- 3 fA base line?
z Add shielded prober but no probes
z Add probes
Debugging A Noisy Probe System
Where to Start?
This is a systematic strategy for focusing in on the sources of noise and error. You can be very
surprised. Some prober cable is not low loss. You may see popping noise due to relaxing of Teflon
fiber in the cables, high current level due to poor dielectric loss, etc. You may find improper triax
adapters, high levels of stray capacitance, etc., as you add prober and probe connections.