Agilent Technologies B2902 Servisní příručka Strana 159

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 326
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 158
Agilent B2900 SCPI Command Reference, Edition 2 4- 37
Subsystem Commands
:HCOPy:SDUMp:DATA?
HCOPy Subsystem
:HCOPy:SDUMp:DATA?
Returns the data of the front panel screen image. The format of the image data is set
by the
:HCOPy:SDUMp:FORMat command.
Syntax :HCOPy:SDUMp:DATA?
Query response The response is a definite length arbitrary binary block.
Example :HCOP:SDUM:DATA?
:HCOPy:SDUMp:FORMat
Sets the format of the image data. The front panel screen image will be created in
the format set by this command. The image data will be returned by the
:HCOPy:SDUMp:DATA? command.
Syntax :HCOPy:SDUMp:FORMat format
HCOPy:SDUMp:FORMat?
Parameter format Format of image data. JPG (default)|BMP|PNG|WMF.
Parameter data type is CPD.
Query response format <newline>
format returns JPG, BMP, PNG, or WMF. Response data type is CRD.
Example :HCOP:SDUM:FORM BMP
:HCOP:SDUM:FORM?