Agilent B2900 SCPI Command Reference, Edition 2 4- 149
Subsystem Commands
:TRACe:TSTamp:FORMat
Parameter operation Statistical operation. MINimum|MAXimum|SDEViation|
PKPK|MEAN (default). Parameter data type is CPD.
operation=MEAN sets the operation for obtaining the mean value.
operation=SDEV sets the operation for obtaining the standard deviation.
operation=PKPK sets the operation for obtaining the peak to peak value.
operation=MIN sets the operation for obtaining the minimum value.
operation=MAX sets the operation for obtaining the maximum value.
Query response operation <newline>
operation returns the present setting of the statistical operation, MEAN, SDEV,
PKPK, MIN, or MAX. Response data type is CRD.
Example :TRAC:STAT:FORM PKPK
:TRAC2:STAT:FORM?
:TRACe:TSTamp:FORMat
Selects the rule for reading the timestamp data in the trace buffer.
Syntax :TRACe[c]:TSTamp:FORMat rule
:TRACe[c]:TSTamp:FORMat?
Parameter rule Rule for reading the timestamp data. DELTa|ABSolute
(default). Parameter data type is CPD.
rule=ABS sets the returned data to the incremental value for the first timestamp
data.
rule=DELT sets the returned data to the incremental value for the previous
timestamp data.
Query response rule <newline>
rule returns the present setting of the rule, DELT or ABS. Response data type is
CRD.
Example :TRAC:TST:FORM DELT
:TRAC2:TST:FORM?