Agilent B2900 SCPI Command Reference, Edition 2 2- 29
Subsystem Command Summary
Controlling Source/Measure Trigger
Controlling Source/Measure Trigger
Figure 2-6 Transient and Acquire Device Actions
NOTE If channels are set as shown below, the device actions start simultaneously.
To synchronize transient actions (source output)
Trigger source is set to the same mode.
Delay time is set to the same value.
Source output ranging mode is set to the fixed mode.
Source wait time control is set to OFF.
Measurement wait time control is set to OFF.
Measurement ranging mode is set to the fixed mode.
TRANsient device action
- for voltage or current output -
TRAN:DELay
ACQuire device action
- for voltage or current measurement -
Trigger
Step value
Step value
Measure Measure
TRAN:DELay
ACQ:DELay ACQ:DELay
Trigger
Trigger
Trigger