3. Check Channels Leakage
Send CLOS(@nn00) to the FET Multiplexer to close channel
00, where nn is card number (typically 01)
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, HI to LO, Channels
Send OPEN (@nn00) to the FET Multiplexer
4. Check Tree Leakage
Send SCAN:PORT ABUS to the FET Multiplexer to enable the
Tree Switches
Send CLOS(@nn00) to the FET Multiplexer to close channel
00 and Tree Switch A, where nn is the card number (typically
01)
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, HI to LO, Tree A
Send OPEN (@nn00) to the FET Multiplexer
Send CLOS(@nn15) to the FET Multiplexer to close channel
15 and Tree Switch B
Send TRIG SGL to DMM
Record the DMM reading in Table 2-1
Positive Polarity, HI to LO, Tree B
Send *RST to the FET Multiplexer
5. Change Polarity
Turn power supply and mainframe power OFF
Connect DMM, power supply, and resistor as shown in
Figure 2-7
Turn power supply and mainframe power ON
Set power supply output to +10 Vdc ± 0.1 Vdc
6. Repeat Steps 2 through 4
Record all results in Table 2-1 as
Negative Polarity, HI to LO
Verification Tests 2-13