Command Reference 5
Programmer’s Guide 131
Spectrogram Marker Audio Alert Limit
[:SENSe]:SPECtrogram:MARKer:ALERt:LIMit <value>
[:SENSe]:SPECtrogram:MARKer:ALERt:LIMit?
Use this command to set the Marker audio alert limit line value.
Example: :SPECtrogram:MARKer:ALERt:LIMit -50 dBm
Key access: Marker > More 1 of 2 > Audio Alert
Spectrogram Start Save
[:SENSe]:SPECtrogram:SAVE:STARt
[:SENSe]:SPECtrogram:SAVE[:STATe]?
Use this command to start to save the spectrogram measurement result.
Example: :SPECtrogram:SAVE:STARt
Key access: Meas > File Logging > Start Save
Spectrogram Stop Save
[:SENSe]:SPECtrogram:SAVE:STOP
[:SENSe]:SPECtrogram:SAVE[:STATe]?
Use this command to stop to save the spectrogram measurement result.
Example: :SPECtrogram:SAVE:STOP
Key access: Meas > File Logging > Stop Save
Spectrogram Time Save State
[:SENSe]:SPECtrogram:SAVE:TIMed[:STATe]
OFF|ON|0|1
[:SENSe]:SPECtrogram:SAVE:TIMed[:STATe]?
Use this command to turn on/off timed save state of spectrogram
measurement.
Example: :SPECtrogram:SAVE:TIMed ON
Key access: Meas > File Logging > Time Setting > Timed Save