Agilent Technologies B2902 Uživatelská příručka Strana 114

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 232
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 113
4- 8 Agilent B2900 User’s Guide, Edition 2
Front Panel Reference
Display and Assist Keys
Single View
Displays the measurement results, source and measurement settings, and channel
status. Most setup parameters can be edited on this display. The lower half displays
the setup parameters selected by Assist keys 2.
Display area 1. Channel number. 1 or 2.
2. Latest measurement data
3. High capacitance (HC) mode indicator
4. Remote sensing (4-wire connection) status indicator
5. Source function. VOLTS or AMPS.
6. Source shape indicator. DC, pulse, sweep, or pulsed sweep. DC does not show
the indicator.
7. Source output value, 5½ digit resolution on B2901A/B2902A, and 6½ digit
resolution on B2911A/B2912A
8. Limit (Compliance) value
9. Measurement speed. AUTO, SHORT, MEDIUM, NORMAL, LONG, or
MANUAL.
10. Aperture time in seconds or PLC (power line cycle, number of power line cycles
per a measurement). Only for the MANUAL speed.
D1. Primary measurement data
D2. Secondary measurement data or limit test result Pass or Fail
1
2
3
4
5
6
7
8
9
10
D1
D2
Zobrazit stránku 113