Commands by Subsystem 5
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 591
:TRIGger:SEQuence:TIMer
(see page 798)
Command Syntax
:TRIGger:SEQuence:TIMer <time_value>
<time_value> ::= time in seconds in NR1 format
The :TRIGger:SEQuence:TIMer command sets the sequencer timer resource
in seconds from 10 ns to 10 s. The timer is used in the reset stage to
determine how long to wait for the trigger to occur before restarting.
Query Syntax
:TRIGger:SEQuence:TIMer?
The :TRIGger:SEQuence:TIMer? query returns current sequencer timer
setting.
Return Format
<time value><NL>
<time_value> ::= time in seconds in NR1 format
See Also "Introduction to :TRIGger Commands" on page 479