Obsolete and Discontinued Commands 7
Agilent InfiniiVision 6000 Series Oscilloscopes Programmer's Guide 715
:EXTernal:PMODe
(see page 798)
Command Syntax
:EXTernal:PMODe <pmode value>
<pmode value> ::= {AUTo | MANual}
The probe sense mode is controlled internally and cannot be set. If a
probe with sense is connected to the specified channel, auto sensing is
enabled; otherwise, the mode is manual.
If the pmode sent matches the oscilloscope's setting, the command will be
accepted. Otherwise, a setting conflict error is generated.
Query Syntax
:EXTernal:PMODe?
The :EXTernal:PMODe? query returns AUT if an autosense probe is
attached and MAN otherwise.
Return Format
<pmode value><NL>
<pmode value> ::= {AUT | MAN}
NOTE
The :EXTernal:PMODe command is an obsolete command provided for compatibility to
previous oscilloscopes.