Agilent Technologies 8753ET Specifikace Strana 12

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 262
  • Tabulka s obsahem
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 11
1-4
8753ES Specifications and Characteristics
Corrected System Performance
Table 1-2 7-mm Device Connector Type
Network Analyzer: 8753ES, Standard and Option 006
Calibration Kit: 85031B (7-mm, 50 )
Cables: 11857D
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 25 ± 5 °C with < 1 °C deviation from cal temp, Isol cal with avg = 8
Description
Typical Specification
30 to 300 kHz 300 kHz to
1.3 GHz
1.3 to 3 GHz 3 to 6 GHz
(Opt 006 only)
Directivity 55 dB 55 dB 51 dB 46 dB
Source Match 55 dB 51 dB 49 dB 43 dB
Load Match 55 dB 55 dB 51 dB 46 dB
Refl. Tracking
Magnitude (db) ±(0.001 + .02/°C) ±(0.001 + .01/°C) ±(0.005 + .02/°C) ±(0.020 + .03/°C)
Phase (deg) ±(0.007 + .05/°C) ±(0.007 + .05/°C) ±(0.033 + .05/°C) ±(0.132 + 0.2/°C)
Trans. Tracking
Magnitude (db) ±(0.009 + .02/°C) ±(0.006 + .01/°C) ±(0.009 + .02/°C) ±(0.021 + .03/°C)
Phase (deg) ±(0.060 + .05/°C) ±(0.040 + .05/°C) ±(0.059 + .05/°C) ±(0.139 + 0.2/°C)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)
Zobrazit stránku 11
1 2 ... 7 8 9 10 11 12 13 14 15 16 17 ... 261 262

Komentáře k této Příručce

Žádné komentáře