8753ES Specifications and Characteristics
Corrected System Performance
Table 1-4 3.5-mm Device Connector Type
Network Analyzer: 8753ES, Standard and Option 006
Calibration Kit: 85033D (3.5-mm, 50 )
Cables: 11857D
Calibration: Full 2-Port
IF BW = 10 Hz, Avg off, Temp = 25 ± 5°C with < 1°C deviation from cal temp, Isol cal with avg = 8
Description
Typical Specification
30 to 300 kHz 300 kHz to
1.3 GHz
1.3 to 3 GHz 3 to 6 GHz
(Opt 006 only)
Directivity 49 dB 46 dB 44 dB 38 dB
Source Match 49 dB 44 dB 41 dB 37 dB
Load Match 49 dB 46 dB 44 dB 38 dB
Refl. Tracking
Magnitude (db) ±(0.001 + .02/°C) ±(0.005 + .01/°C) ±(0.007 + .02/°C) ±(0.009 + .03/°C)
Phase (deg) ±(0.007 + .05/°C) ±(0.033 + .05/°C) ±(0.046 + .05/°C) ±(0.059 + 0.2/°C)
Trans. Tracking
Magnitude (db) ±(0.019 + .02/°C) ±(0.014 + .01/°C) ±(0.022 + .02/°C) ±(0.048 + .03/°C)
Phase (deg) ±(0.124 + .05/°C) ±(0.092 + .05/°C) ±(0.145 + .05/°C) ±(0.317 + 0.2/°C)
Transmission Uncertainty (Specification)
Reflection Uncertainty (Specification)