Chapter 1 1-19
Alphabetical Command Reference
BWLIMMAX
BWLIMMAX
Syntax
BWLIMMAX<num>[HZ|KHZ|MHZ|GHZ];
or
BWLIMMAX?;
Description
Front Panel Equivalents
BWLIMMIN
Syntax
BWLIMMIN<num>[HZ|KHZ|MHZ|GHZ];
or
BWLIMMIN?;
Description
Front Panel Equivalents
Command Description Range Query Response
BWLIMMAX Enters the maximum bandwidth value. If the
measured bandwidth is greater than this
value, the filter fails the bandwidth test.
stimulus range
1
1. Refer to “Preset State and Memory Allocation” in your analyzer’s reference guide.
<num><
L
F
>
Command Hardkey Softkey
BWLIMMAX
Command Description Range Query Response
BWLIMMIN Enters the minimum bandwidth value. If
the measured bandwidth is less than this
value, the filter fails the bandwidth test.
stimulus range
1
1. Refer to “Preset State and Memory Allocation” in your analyzer’s reference guide.
<num><
L
F
>
Command Hardkey Softkey
BWLIMMIN
System
MAXIMUM BANDWIDTH
System
MINIMUM BANDWIDTH