Agilent Technologies EXG Uživatelská příručka Strana 127

  • Stažení
  • Přidat do mých příruček
  • Tisk
  • Strana
    / 195
  • Tabulka s obsahem
  • ŘEŠENÍ PROBLÉMŮ
  • KNIHY
  • Hodnocené. / 5. Na základě hodnocení zákazníků
Zobrazit stránku 126
Care and Cleaning
7-2 Maintaining the Cardiograph
Cleaning the Electrodes and Cables
Clean the electrodes and patient cables with a soft cloth moistened with a rec-
ommended disinfectant or cleaning agent from the following list:
l Cetylcide® (may discolor cable)
l Cidex®
l Lysol® Disinfectant
l Lysol® Deodorizing Cleaner (may discolor cable)
l Dial® Liquid Antibacterial Soap
l Ammonia
l 409® (may discolor cable)
l 10% solution of Clorox® in water (may discolor cable)
l Murphy® Household Cleaner, or
l Ves-phene II®.
Wring any excess moisture from the cloth before cleaning.
CAUTION Do not clean the patient cable with alcohol. Alcohol can cause the plastic to
become brittle and may cause the cable to fail prematurely.
Do not autoclave the cable or use ultrasonic cleaners.
Do not immerse the patient cable.
Do not use abrasive materials to clean metal surfaces—scratches on them can
cause artifacts on the ECG.
Do not wet the connectors, especially the 15-pin connector.
Zobrazit stránku 126
1 2 ... 122 123 124 125 126 127 128 129 130 131 132 ... 194 195

Komentáře k této Příručce

Žádné komentáře